4-6 | Heated Acid Gas Sample Probe (Large Diameter)
Changing Out Replaceable Parts
This section describes replacing parts in the HAG Sample Probe, as per the
Preventive Maintenance Schedule.
Performing preventive maintenance on the analyzer/HAG Probe
requires working with the analyzer’s System 200 Configurator
Software. Familiarize yourself with the Software before working
on the analyzer and HAG Probe.
Replacing HAG Probe Parts
This procedure discusses maintenance procedures to change out parts in
the HAG Probe. The Sample/Vent Isolation valves and (optional) Double
Block valves in the HAG Probe do not require regular maintenance.
It is necessary to change out the o-rings, Hi-Flow Membrane, Fiber Filter
Element, and Sintered Disk:
• Every six months.
• If evidence of contamination and/or leaks is present.
It is necessary to change out the Sample/Vent Isolation valves or (optional)
Double Block valves immediately if:
• Leaks are originating from any of these valves.
IMPORTANT
Be sure to have the required parts (o-rings, Hi-Flow Membrane, Fiber
Filter Element, Sintered Disk, and/or Sample/Vent Isolation valve or
(optional) Double Block valve kits on-site prior to disassembling the
Sample Probe valve assembly.
NOTE
See Chapter 5 for
spare parts ordering numbers.
See Chapter 5 for
extended maintenance spare
parts ordering numbers.
NOTE
Содержание ASR900
Страница 16: ...1 4 Heated Acid Gas Sample Probe Large Diameter Figure 1 2 HAG Sample Probe assembly drawing...
Страница 25: ...Specifications 2 5...
Страница 26: ...2 6 Heated Acid Gas Sample Probe Large Diameter...
Страница 27: ...Specifications 2 7...
Страница 28: ...2 8 Heated Acid Gas Sample Probe Large Diameter...
Страница 29: ...Specifications 2 9...
Страница 52: ...3 22 Heated Acid Gas Sample Probe Large Diameter This page intentionally left blank...
Страница 59: ...Maintenance 4 7 Figure 4 1 Large Diameter Heated Acid Gas Sample Probe Assembly...
Страница 80: ...4 28 Heated Acid Gas Sample Probe Large Diameter This page intentionally left blank...
Страница 86: ...6 2 Heated Acid Gas Sample Probe Large Diameter Heated Acid Gas Sample Probe Assembly 100 2196...
Страница 88: ...6 4 Heated Acid Gas Sample Probe Large Diameter Double Block Valve Assembly 100 2393 SEE NOTES 1 2...
Страница 89: ...Drawings 6 5 Fiber Filter Assembly 100 2194...
Страница 90: ...6 6 Heated Acid Gas Sample Probe Large Diameter This page intentionally left blank...
Страница 92: ...S 2 Heated Acid Gas Sample Probe Large Diameter This page intentionally left blank...