Maintenance | 4-3
Frequency
Task
Other
Extended Maintenance:
Probe Heater
The Heater Assembly temperature zone circuitry does not require
regular maintenance. However, if problems arise with the Heater
(Probe-related alarms are active), determine the origin of the problem
and immediately replace the faulty part(s). These parts can include the
Over-Temperature Switch, Heater Cartridge, and RTD.
Refer to the “Maintenance and Troubleshooting” chapter (see section
“Host Controller Board Error Conditions and Corrective Action”) in the
analyzer manual for information on specific alarms. See “Replacing
Parts in the Probe Heater Assembly” in this chapter.
IMPORTANT
Be sure to have on-site the proper replacement parts prior to
disassembling the Heater Assembly. Refer to the “Service and
Parts” chapter for the part numbers.
Other
Flamepath Gap
During each Probe Heater Assembly maintenance, use a feeler gauge
to check the flamepath gap between the Heater Body and the Heater
Cover (see Figure 4-4 for location). The gap must not exceed 0.04 mm;
if the gap exceeds this value, contact AMETEK for advice. See
“Examining and Caring For the Flamepaths” in this chapter for more
information.
Содержание ASR900
Страница 16: ...1 4 Heated Acid Gas Sample Probe Large Diameter Figure 1 2 HAG Sample Probe assembly drawing...
Страница 25: ...Specifications 2 5...
Страница 26: ...2 6 Heated Acid Gas Sample Probe Large Diameter...
Страница 27: ...Specifications 2 7...
Страница 28: ...2 8 Heated Acid Gas Sample Probe Large Diameter...
Страница 29: ...Specifications 2 9...
Страница 52: ...3 22 Heated Acid Gas Sample Probe Large Diameter This page intentionally left blank...
Страница 59: ...Maintenance 4 7 Figure 4 1 Large Diameter Heated Acid Gas Sample Probe Assembly...
Страница 80: ...4 28 Heated Acid Gas Sample Probe Large Diameter This page intentionally left blank...
Страница 86: ...6 2 Heated Acid Gas Sample Probe Large Diameter Heated Acid Gas Sample Probe Assembly 100 2196...
Страница 88: ...6 4 Heated Acid Gas Sample Probe Large Diameter Double Block Valve Assembly 100 2393 SEE NOTES 1 2...
Страница 89: ...Drawings 6 5 Fiber Filter Assembly 100 2194...
Страница 90: ...6 6 Heated Acid Gas Sample Probe Large Diameter This page intentionally left blank...
Страница 92: ...S 2 Heated Acid Gas Sample Probe Large Diameter This page intentionally left blank...