Electromagnetic Compatibility (EMC)
Read and follow the recommendations in this section to avoid performance variations or dam-
age to the internal circuits of this equipment when installed in harsh electrical environments.
The various configurations of analyzers that use Heated Acid Gas Sample Probes, and ASR900
Heaters used on Heated Acid Gas Sample Probes, should not produce, or fall victim to, electro-
magnetic disturbances as specified in the European Union’s EMC Directive. Strict compliance
to the EMC Directive requires that certain installation techniques and wiring practices are used
to prevent or minimize erratic behavior of the Analyzer or its electronic neighbors. Below are
examples of the techniques and wiring practices to be followed.
In meeting the EMC requirements, the various Analyzer configurations described in this manual
rely heavily on the use of metallic shielded cables used to connect to the customer’s equip-
ment and power. Foil and braid shielded I/O and DC power cables are recommended for use in
otherwise unprotected situations. In addition, hard conduit, flexible conduit, and armor around
non-shielded wiring also provides excellent control of radio frequency disturbances. However,
use of these shielding techniques is effective only when the shielding element is connected to
the equipment chassis/earth ground at both ends of the cable run. This may cause ground loop
problems in some cases. These should be treated on a case-by-case basis. Disconnecting one
shield ground may not provide sufficient protection depending on the electronic environment.
Connecting one shield ground via a 0.1 microfarad ceramic capacitor is a technique allowing
high frequency shield bonding while avoiding the AC-ground metal connection. In the case of
shielded cables the drain wire or braid connection must be kept short. A two-inch connection
distance between the shield’s end and the nearest grounded chassis point, ground bar or ter-
minal is highly recommended. An even greater degree of shield performance can be achieved
by using metallic glands for shielded cable entry into metal Electronics Enclosures. Expose
enough of the braid/foil/drain where it passes through the gland so that the shield materials can
be wrapped backwards onto the cable jacket and captured inside the gland, and tightened up
against the metal interior.
Inductive loads connected to the low voltage “Alarm Contacts” are not recommended. However,
if this becomes a necessity, follow proper techniques and wiring practices. Install an appropriate
transient voltage suppression device (low voltage MOV, “Transzorb,” or R/C) as close as possible
to the inductive device to reduce the generation of transients. Do not run this type of signal wir-
ing along with other I/O or DC in the same-shielded cable. Inductive load wiring must be sepa-
rated from other circuits in conduit by using an additional cable shield on the offending cable.
In general, for optimum protection against high frequency transients and other disturbances, do
not allow installation of this Analyzer where its unshielded I/O and DC circuits are physically
mixed with AC mains or any other circuit that could induce transients into the Analyzer or the
overall system. Examples of electrical events and devices known for the generation of harmful
electromagnetic disturbances include motors, capacitor bank switching, storm related transients,
RF welding equipment, static, and walkie-talkies.
!
CAUTION
Содержание ASR900
Страница 16: ...1 4 Heated Acid Gas Sample Probe Large Diameter Figure 1 2 HAG Sample Probe assembly drawing...
Страница 25: ...Specifications 2 5...
Страница 26: ...2 6 Heated Acid Gas Sample Probe Large Diameter...
Страница 27: ...Specifications 2 7...
Страница 28: ...2 8 Heated Acid Gas Sample Probe Large Diameter...
Страница 29: ...Specifications 2 9...
Страница 52: ...3 22 Heated Acid Gas Sample Probe Large Diameter This page intentionally left blank...
Страница 59: ...Maintenance 4 7 Figure 4 1 Large Diameter Heated Acid Gas Sample Probe Assembly...
Страница 80: ...4 28 Heated Acid Gas Sample Probe Large Diameter This page intentionally left blank...
Страница 86: ...6 2 Heated Acid Gas Sample Probe Large Diameter Heated Acid Gas Sample Probe Assembly 100 2196...
Страница 88: ...6 4 Heated Acid Gas Sample Probe Large Diameter Double Block Valve Assembly 100 2393 SEE NOTES 1 2...
Страница 89: ...Drawings 6 5 Fiber Filter Assembly 100 2194...
Страница 90: ...6 6 Heated Acid Gas Sample Probe Large Diameter This page intentionally left blank...
Страница 92: ...S 2 Heated Acid Gas Sample Probe Large Diameter This page intentionally left blank...