Chapter 9: Additional Features, Page 92
M o d e l 6 8 2 T - H P U S E R M A N U A L A T E X Z o n e 2
Transmission Parameters:
Bits: 8
Parity: Even
Stop bits: 1
Modbus
Word size &
content:
Each Modbus Word is 16 bits, consisting of 2 Bytes. The Byte order in each
Word is Big Endian (MSB first).
Floating point values are single precision IEEE 754 standard. Floating point
values require 2 contiguous Word's.
682T-HP MODBUS WORD ASSIGNMENTS
Address Description
Bit
HEX
Value
O40001
Status Word:RUN BIT = 1 when system is in RUN
0
0x0001
O40001
Status Word:MODE LOW BIT = set by Control Word when system is in RUN
1
0x0002
O40001
Status Word:MODE HIGH BIT = set by Control Word when system is in RUN
2
0x0004
O40001
Status Word:ALARM BIT = 1 when system has NO alarm
3
0x0008
O40001
Status Word:PURGE OK BIT = 1 when purge is OK
4
0x0010
O40001
Status Word:LOW DENSITY BIT = 1 when density is below the configured limit
5
0x0020
O40001
Status Word:VERIFY BIT = 1 when system is performing a verify operation
6
0x0040
O40001
Status Word:DATA UPDATE BIT = 1 when results have been recently updated (5
second pulse)
7
0x0080
O40001
Status Word:OUT OF BOUNDS BIT = 1 when results are outside configured
bounds
8
0x0100
O40001
Status Word:RESERVED BIT = reserved for future use
9
0x0200
O40001
Status Word:RESERVED BIT = reserved for future use
10
0x0400
O40001
Status Word:RESERVED BIT = reserved for future use
11
0x0800
O40001
Status Word:RESERVED BIT = reserved for future use
12
0x1000
O40001
Status Word:RESERVED BIT = reserved for future use
13
0x2000
O40001
Status Word:RESERVED BIT = reserved for future use
14
0x4000
O40001
Status Word:RESERVED BIT = reserved for future use
15
0x8000
O40002
Control Word:RUN BIT set = 1 when system RUN desired, set = 0 to stop system
0
0x0001
O40002
Control Word:MODE LOW BIT set = 1 for CalSet1 or CalSet3, set = 0 for CalSet0
or CalSet2
1
0x0002
O40002
Control Word:MODE HIGH BIT set = 1 for CalSet2 or CalSet3, set = 0 for CalSet0
or CalSet1
2
0x0004
O40002
Control Word:RESERVED BIT = reserved for future use
3
0x0008
O40002
Control Word:RESERVED BIT = reserved for future use
4
0x0010
O40002
Control Word:RESERVED BIT = reserved for future use
5
0x0020
O40002
Control Word:RESERVED BIT = reserved for future use
6
0x0040
O40002
Control Word:RESERVED BIT = reserved for future use
7
0x0080
O40002
Control Word:RESERVED BIT = reserved for future use
8
0x0100
O40002
Control Word:RESERVED BIT = reserved for future use
9
0x0200
O40002
Control Word:RESERVED BIT = reserved for future use
10
0x0400
O40002
Control Word:RESERVED BIT = reserved for future use
11
0x0800
Содержание 682T-HP
Страница 1: ...Model 682T HP Sulfur Analyzer ATEX Zone 2 pzc User Manual 9000 223 VE REV C...
Страница 25: ...Chapter 3 Operation Page 18 M o d e l 6 8 2 T H P U S E R M A N U A L A T E X Z o n e 2...
Страница 56: ...Chapter 5 Hardware Details Page 49 M o d e l 6 8 2 T H P U S E R M A N U A L A T E X Z o n e 2...
Страница 66: ...Chapter 5 Hardware Details Page 59 M o d e l 6 8 2 T H P U S E R M A N U A L A T E X Z o n e 2...
Страница 67: ...Chapter 5 Hardware Details Page 60 M o d e l 6 8 2 T H P U S E R M A N U A L A T E X Z o n e 2...
Страница 70: ...Chapter 5 Hardware Details Page 63 M o d e l 6 8 2 T H P U S E R M A N U A L A T E X Z o n e 2...
Страница 72: ...Chapter 5 Hardware Details Page 65 M o d e l 6 8 2 T H P U S E R M A N U A L A T E X Z o n e 2...
Страница 75: ...Chapter 5 Hardware Details Page 68 M o d e l 6 8 2 T H P U S E R M A N U A L A T E X Z o n e 2...
Страница 76: ...Chapter 5 Hardware Details Page 69 M o d e l 6 8 2 T H P U S E R M A N U A L A T E X Z o n e 2...