Fixed Spot Thermometers
User Guide
SPOT Actuator
SPOT
SPOT
Type
Select the type of SPOT thermometer from the drop-down
list. Note that the ‘Scan’ option on the I/O menu of the SPOT
thermometer must also be set to enable correct scan timing.
4mA
and
20mA
The actuator reads the 4 to 20 mA output from the SPOT
thermometer, so the 4 mA and 20 mA equivalent temperatures
must match those of the SPOT for correct interpretation of
temperature. 200 °C and 700 °C are standard limits for the
AL EQS thermometer. However it is possible to configure both
SPOT and actuator to use a different range if required for
interaction with other connecting devices.
IP
The IP address of the SPOT AL EQS Thermometer. If this is set
correctly, the SPOT Webserver can be accessed from the main
page of the Actuator Webserver. It is not required for any other
actuator functionality.
4.5.3 Display Settings
[0 ]C
(0 = Auto)
You can specify the upper and lower limits of temperature to
be displayed in the profile. If you leave these as 0, the profile
is scaled automatically to cover the maximum and minimum
temperatures being measured.
Text input
boxes
You can give a label to either end of the profile (Maximum 8
characters)
Click on the arrow to switch the Left-Right direction to match the
orientation of your display to that of your Actuator.