37
11.1.3 IP Rating
The unit has an
IP21
rating.
2
= Protected against solid objects greater than 0.492" (12.5mm)
1
= Protected from vertically dripping water
11.1.4 Alarm port connection details
Type:
Volt free/dry changeover contacts
Minimum applied Voltage:
34Vdc/24Vac
Maximum applied current:
1 Amp
Details of this connection shall be provided upon request, contact
AmeriWater
or your Healthcare provider.
11.1.5 Water quality and performance
Max drain flow-
rate @ 10
o
C
USgals/min
Permeate
output @ 10
o
C
USgals/min
Recovery
%
Output water quality
0.29
0.4
60%
Will meet the requirements of
current AAMI/ANSI/ISO:13959
standard for
“water for
hemodialysis and related
therapies”
11.1.6 Feed water requirements
Pre-filtration
Filtered to 5 microns
Total Hardness
Maximum 400 ppm as CaCO
3
Temperature
34 – 95
o
F (1-35
o
C)
Chlorine (Total)
<0.1 ppm free Cl
2
Total dissolved solids (max)
1500 mg/l
Fouling index
<5
Feed water pressure
2-6 bar (30-90 psi)
Feed water flowrate
1-1.5USgals/min (3.8-5.7ltrs/min)
11.1.7 Water services connections
Connection
Description
Size
Type
Drain
Unit waste water-out
8mm
Push fit
In
Feedwater supply-in
8mm
Push fit
Out
Permeate-out
8mm
Push fit
Return
Returned permeate-in
8mm
Push fit
Acid
Chemical disinfectant-in
4mm
Push fit