3.4.2  Residual dangers 

Warning: 

Electrical Shock

Do Not 

remove the side covers or touch or interfere with any of the electrical components.

Mechanical force. 

Some parts of the system can be under pressure of up to 215 psi (15 bar). Do Not remove the side covers 
and interfere with the water pipework or any water connections, as you could risk the possibility of injury 
to yourself or others or cause damage to surrounding equipment. 

Hot Surfaces. 

During the heat disinfection cycle the external  distribution pipework and connector to 

the dialysis machine will become hot to touch, avoid touching these components and warn others.

3.4.3  Handling 

Warning: 

 The unit has been specifically designed as a stationary device. Should the unit need to be relocated it 
must be decommissioned before moving. Refer to 

Section 4.14 “Decommissioning for relocation 

procedure” 

in the 

Installation & Commissioning Guide.

 The dry weight of the unit is 86 pounds.  A 

transport cart or other suitable device should be used when moving the unit. Precautions should be 
taken to secure the unit from toppling during relocation. 

Do Not

 pick the unit up by the side covers; they are not designed to take the weight of the unit. Support 

the unit by holding the underside of the chassis while steadying the body. 

When relocating the unit always ensure the unit has performed its full drain down routine. To do this switch 
the unit on, isolate the incoming mains water supply, press 

“START”

 and run the unit until it stops due to 

‘Low Tank Level’

 then switch off.  

Do Not

 move the unit while it is still in operation. 

3.5  Bringing the unit to an immediate 

STOP

If you need to stop the unit immediately at any time, simply press the black rocker switch 
on the back of the unit to the 

“O”

 position. Then as a precaution, turn off the water supply 

at source. Alternatively press the red 

“STOP”

 button on the touch-screen twice.  

Unless safe to do so

 Do Not

 restart the  unit until you have spoken to 

AmeriWater

 or 

your  Healthcare  provider  for  advice  or  assistance,  contact  telephone  numbers can  be 
found in 

Section 2.

Warning:  

The supply mains isolation switch is located at the rear of the unit ensure that it is accessible at all times 
and remains clear from obstruction. 

Ensure that the wall plug is accessible at all times as this can be used as a method of isolation. 

If the unit was stopped due to a leak, before reinstating the power to the unit carry out an electrical safety 
inspection as detailed in 

Section 9.1.1

 of the 

“Service & Maintenance”

 manual. 

Содержание CENTURION 1500+

Страница 1: ...CENTURION 1500 OPERATING MANUAL P N 98 2010 REV C...

Страница 2: ......

Страница 3: ...Revision C December 2017 CENTURION 1500 UF OPERATING MANUAL For L998374 5 DOCUMENT HISTORY Revision Date of Change A January 2013 B January 2016 C December 2017...

Страница 4: ...ures and process of operation 12 4 5 Explanation of rear connections 13 5 0 OPERATING THE TOUCHSCREEN 14 5 1 Explanation of Buttons 14 5 1 1 Using the screen 14 5 1 2 MENU button 14 5 1 3 START button...

Страница 5: ...6 Feed water requirements 37 11 1 7 Water services connections 37 11 1 8 Raw water break tank 38 11 1 9 Weights and dimensions 38 11 1 10 Main PCB battery 38 11 1 11 USB mass storage device 38 11 2 En...

Страница 6: ...r a qualified technician to maintain and service the unit Details on how to install commission the Centurion 1500 can be found in the Installation Commissioning Guide Installation of the unit would al...

Страница 7: ...ly contact your Healthcare provider or AmeriWater or refer to the relevant accompanying manuals and guides detailed in Section 1 0 PREFACE AmeriWater Tel No 800 535 5585 Or your local authorized Ameri...

Страница 8: ...formation that can be used to help explain a command or action or give guidance 3 2 Explanation of labels There are a number of labels applied to the outside of the Centurion 1500 unit which identify...

Страница 9: ...th this label during heat disinfection or take suitable measures to protect yourself from the heat Do Not Sit Under no circumstances should the unit be sat on or objects placed on the top sloping cove...

Страница 10: ...ean This label identifies the position of the key When approved trained to do so always refer to the Chemical cleaning and heat disinfection instructions if unsure about when and how to use this key E...

Страница 11: ...tion 11 1 1 The Refer to the manual symbol will direct your Healthcare provider installer to the Installation Commissioning guide This label provides details of all the water connections If more detai...

Страница 12: ...iately take the necessary precautions if the spillage contains bodily fluids Refer to Section 8 3 Cleaning external surfaces when dealing with spillages If the unit develops a leak follow the procedur...

Страница 13: ...as part of a Medical Electrical ME system The unit should not be stacked on top off or placed directly adjacent to other electrical electronic equipment The Centurion 1500 water purifier works on the...

Страница 14: ...ot designed to take the weight of the unit Support the unit by holding the underside of the chassis while steadying the body When relocating the unit always ensure the unit has performed its full drai...

Страница 15: ...ned and operated according to the instructions contained within this Operating manual and supporting documents AmeriWater Healthcare provider will not accept any responsibility for any damage or injur...

Страница 16: ...t The unit is fitted with an internal pump that pressurises the water supply and forces it through a membrane which then separates out all of the impurities from the feedwater Purified water termed pe...

Страница 17: ...tank and boosts the water pressure to the RO membrane There is a pressure sensor situated immediately after the pump which will display a warning message on the touch screen should the maximum system...

Страница 18: ...13 4 5 Explanation of rear connections Key Switch Alarm Output USB port On Off switch Fuses Power lead socket Acid cleaner inlet Water In Water Out Water Return Drain Cooling Fan...

Страница 19: ...ion simply touch the screen on the appropriate button once with your finger Note Do Not use sharp or pointed implements such as pens pencils etc to operate the screen as this will damage the sensitive...

Страница 20: ...tch to the I position When powered the unit s HMI screen will illuminate and display both text and a visual mimic of the units operation The touch screen will initially display Intialization for a few...

Страница 21: ...e STOP button twice from the processing screen will stop the unit immediately terminating the flush Refer to Section 3 5 Warning The supply mains isolation switch is located at the rear of the unit en...

Страница 22: ...measured in microsiemens cm or S cm Temperature This tells you the temperature of the water being produced and is displayed in 0 F Pump Pressure The unit contains a pump to process the water and this...

Страница 23: ...ch are only provided for use by your Healthcare provider The various sub menu options can be selected by pressing the buttons followed by ENTER 6 1 1 Set Time and Date From the USER MENU and by using...

Страница 24: ...p stating Done and after a few seconds the screen will revert back to the Time Date Menu To select Set Date from the menu use the buttons and press ENTER The date will be displayed and the cursor will...

Страница 25: ...ecomes due 6 1 3 Product Information This screen shows the product details and software version numbers and is for information only 12 04 12 6 MONTH INTERVAL Pre Filter Date RESET BACK Product Informa...

Страница 26: ...ettings Menus are only accessible for use by trained and authorized personnel If any data held within these menus is altered the performance safety of the unit and or water quality may be seriously af...

Страница 27: ...to carry out cleaning disinfection of the unit When approved to do so ensure that the procedures detailed within this document are followed Warning Chemical cleaning of the unit MUST only be carried b...

Страница 28: ...lt in a deterioration of the water quality provided by the unit making it unsuitable for hemodialysis During heat disinfection of the unit there is a potential for water exceeding 1760F to be discharg...

Страница 29: ...ed water output Min 0 40 USgals min 500F Monthly Check to be carried out by Healthcare provider Hot water disinfection Min temp 1760F Min hold time 10 mins Once per week min Record of last heat disinf...

Страница 30: ...r leaks in or around the unit Verify the conductivity on the Centurion display is below the alarm set point Note Do Not start dialysing if the unit appears to have a fault or the water quality is in d...

Страница 31: ...0F On starting the unit up after standing run for 10 minutes disconnected from the dialysis machine and carryout a heat disinfection Your Healthcare provider will advise you on the best method of main...

Страница 32: ...follow the instructions in Section 3 5 to shut the unit down quickly Once the unit has been safely shut down and the water supply turned off make a note of any messages that were displayed on the scre...

Страница 33: ...e water supply 1 DO NOT try to run the unit in this condition contact your Healthcare provider for assistance Low Pressure The unit has detected insufficient pressure to operate 1 Check that the feedw...

Страница 34: ...d not tilted over 1 To prevent the possibility of water leakage due to the internal tank overflowing for safety it will have stopped running 2 No other warning messages are displayed and the unit appe...

Страница 35: ...ngineer has switched the unit off many times to repair a fault 1 Ensure that the mains electrical lead is secure and fixed 2 Check the condition of your circuit breaker and any other household applian...

Страница 36: ...our Healthcare provider for advice Poor Water Quality The unit has detected that the quality of the permeate water is above 200 S cm Refer to Service Due message previously detailed Note Should the pe...

Страница 37: ...entilation fan is not obstructed or the unit covered by anything and placed in a well ventilated area at ambient temperature away from any heat sources 1 Advise your Healthcare Provider immediately do...

Страница 38: ...r trained qualified electrical engineer that is familiar with this equipment There is insufficient flow to run the dialysis machine 1 A fault has occurred with the unit that has either stopped the uni...

Страница 39: ...ossible causes Checks Proposed actions Bacterial count and or Endotoxin levels in dialysis water exceed recommended guidelines 1 Fault with Heat disinfection cycle 2 Period between Heat disinfections...

Страница 40: ...efer to section 8 2 Pre dialysis checks to assess performance of RO membrane 3 Repeat samples to confirm levels 4 Check pre treatment is functioning correctly 5 Refer to Section 8 1 1 for detail of ch...

Страница 41: ...ective earth For permanent installations the mains supply must be provided with a Branch Circuit Breaker refer to Section 4 6 2 of the Installation Commissioning Guide for details of rating and specif...

Страница 42: ...utput water quality 0 29 0 4 60 Will meet the requirements of current AAMI ANSI ISO 13959 standard for water for hemodialysis and related therapies 11 1 6 Feed water requirements Pre filtration Filter...

Страница 43: ...pe Lithium CR2032 11 1 11USB mass storage device Specification FAT 16 formatted USB memory stick Memory size Must be less than 2GB 11 2 Environmental data Parameter Normal Operation Storage Transport...

Страница 44: ...ssions CISPR 11 Group 1 The Centurion 1500 uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equip...

Страница 45: ...l commercial or hospital environment Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth A A Mains power quality should be that of a typical commercial or hospital environment Voltage dips...

Страница 46: ...ers m Field strength from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance lever in each frequency range b Interference may occur in the vicin...

Страница 47: ...rs m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufactur...

Страница 48: ...he device bears the following UL mark CLASSIFIED BY UNDERWRITERS LABORATORIES INC WITH RESPECT TO ELECTRIC SHOCK FIRE MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH ANSI AAMI ES60601 1 2005 3rd ed and CAN...

Страница 49: ...thout deforming it or restricting flow Push the tube into the connector until the tube stops Ensure the tube is pushed in as far as the tube stop pull on the tube slightly to check that it is secured...

Страница 50: ...nent from the unit must be in accordance with local requirements in your province or state for the disposal of electrical waste E Waste Your Healthcare provider will be responsible for the disposal of...

Страница 51: ...46 12 3 Flow schematic...

Страница 52: ...47 12 4 Certificate of conformity...

Отзывы: