36
11.0
Technical specification
11.1 Services/ connections and physical properties
11.1.1 Electrical specifications/connections
Mains supply
Electrical
supply
Operation
Max Power
consumption
(Watts)
Single
phase
115V 60Hz
plus earth
Standby
10
Normal
operation
160
During heat
disinfection
1000
Warning:
To avoid risk of electric shock, this equipment must be connected to a supply main
with protective earth.
For permanent installations the mains supply must be provided with a Branch Circuit Breaker,
refer to
Section 4.6.2
of the
Installation & Commissioning Guide
for details of rating and
specification of Branch Circuit Breaker.`
11.1.2 Fuse rating/type
External:
(Located at the rear of the unit)
Type:
2 off 12 Amp
T12AH115V
- 5mm x 20mm,
Ceramic, time delay
Internal:
(Located in main electrical tray)
Type:
2 off 5 Amp
T5AH115V
– 5mm x 20mm, Ceramic, time delay
Note:
Fuses must only be replaced with those approved and supplied by the
manufacturer, refer to Service and Maintenance Manual for details.