15
Status
Break
Tank
Inlet
Solenoid
Stop
Button
RO Pump
Function
Drain
Solenoid
RO Module
Permeate
Solenoid
STOP
RO
PROCESSING
15.09.09 12:36
5.1.3 “START” button
Pressing the
“START”
button the unit will go into
“PROCESSING”
mode and start
to produce purified water and the screen below will be displayed.
Refer to
Section 5.3
for explanation of the mimic screen.
5.2
Operation
5.2.1
Initial Start up Procedure
Carryout the Pre-Dialysis checks listed in
Section 8.2.
Once you have completed the above checks switch on the unit using the,
I/O
(on/off)
switch at the rear of the unit. Switch to the
"I"
position.
When powered the unit’s HMI screen will illuminate and display both text and a visual
mimic of the units operation.
The touch screen will initially display
”Intialization”
for a few seconds before
showing the
Power On
screen, which displays the current date, time, model type and
any active messages.
During
“Intialization”
, two “bleeps” will sound indicating that the audible alarm is
functioning correctly.
5.2.2 Starting the unit
To Start the unit select the “
START”
button, the screen will then change to the
“PROCESSING”
screen which will display a mimic of the unit’s operation.
The unit will then carry out a high flow flush to drain, followed by an internal permeate
rinse. After 90 seconds the permeate will automatically flow to the dialysis machine,
if connected.