8.5
Image Quality And Brightness
Possible Problem
Possible Cause
Solution
No video image
aBox ™ Duodeno or ancillary
equipment not switched on.
Switch aBox™ Duodeno and
ancillary equipment on.
Endoscope connector is not
properly connected to the
aBox™ Duodeno.
Connect the endoscope con-
nector properly to the aBox™
Duodeno.
LED´s switched OFF.
Switch LED´s ON.
Medical grade monitor
defective
Contact support for the parti
-
cular device.
Medical grade monitor not
or not properly connected.
Connect the medical grade
monitor properly.
aScope™ Duodeno defect
Withdraw the aScope™
Duodeno carefully and
connect a new aScope™
Duodeno
aBox™ Duodeno defective
Contact support
Image suddenly darkens
Camera or illumination
failure
Withdraw the aScope™
Duodeno and connect a new
aScope™ Duodeno
Cooling malfunction
Connect a new bottle of
sterile water and/or insert the
process water tubing so that
it reaches the bottom of the
bottle if necessary or change
suction canister if necessary.
Blurry image
Objective lens is dirty
Rinse the objective lens
Water drops on the lens
Insufflate to remove water
drops from the lens
Image settings incorrect
aBox™ Duodeno
See aBox™ Duodeno IFU
Dark or over-illuminated
image
Image settings incorrect
aBox™ Duodeno
See aBox™ Duodeno IFU
aScope™ Duodeno defective
Connect a new aScope™
Duodeno
The color tone of the endo
-
scopic image is unusual
aScope™ Duodeno defective
Connect a new aScope™
Duodeno
Picture is frozen
Freeze function is ON
Release freeze function
aScope™ Duodeno defective
Connect a new aScope™
Duodeno
aBox™ Duodeno defective
Contact support