Ancillary Equipment

• 

If combinations of ancillary equiment other than those described in this IFU are used, full 

responsability is assumed by the medical treatment facility.

• 

A defective vacuum pump could lead to a loss of cooling and increased temperature at 

the tip of the endoscope. Make sure to have another vacuum source available. 

 

Operation of the Endoscope

• 

Never  perform  angulation  control  forcibly  or  abruptly.  Never  forcefully  pull,  twist,  or 

rotate  the  angulated  bending  section.  Patient  injury,  bleeding,  and/or  perforation 

may result. It may also become impossible to straighten the bending section during an             

examination.  Never  operate  the  bending  section,  feed  air  or  perform  suction,  insert 

or withdraw the endoscope’s insertion section, or use endoscopic accessories without 

viewing  the  live  endoscopic  image  or  when  the  image  is  frozen  or  magnified.  Patient 

injury, bleeding, and/or perforation may result. 

• 

Never insert, withdraw or operate the insertion section abruptly or with excessive force. 

Patient injury, bleeding, and/or perforation may result.

• 

Never insert or withdraw the endoscope’s insertion section while the bending section is 

bent or locked in position. Patient injury, bleeding, and/or perforation may result.

• 

Never  insert  or  withdraw  the  endoscope  while  the  elevator  is  raised,  or  when                           

endoscopic accessory extends from the distal end of the endoscope. Otherwise, patient 

injury, bleeding, and/or perforation can result.

• 

Firmly connect the suction tube from the suction pump to the suction connector on the 

endoscope. If the suction tube is not attached properly, debris may drip from the tube 

and can pose an infection control risk, cause equipment damage, and/or reduce suction 

capability.

• 

If the sterile water level in the water bottle is too low, replace the bottle with a new one. 

An empty bottle can cause loss of rinsing and cooling function. Always keep a new bottle  

of sterile water ready for this case.

• 

If blood unexpectedly adheres to the surface of the insertion section of the withdrawn 

endoscope, carefully check the condition of the patient.

• 

Do not use the endoscope if any parts of the endoscope or endoscopic accessory fall off 

inside the patient body due to equipment damage or failure. Stop using the endoscope 

immediately and retrieve all parts in an appropriate way, otherwise serious patient injury 

may occur.

Image

Always check whether the image on the screen is a live image or a recorded image and verify 

that the orientation of the image is as expected. Do not use this device if the live image cannot 

be observed, otherwise patient injury may occur.

• 

Always  set  the  intensity  of  electrosurgical  instruments  to  the  lowest  required  value 

to avoid image disturbance. Do not increase the level of intensity above the standard 

ERCP settings. Always limit the continuous activation time of electrical instruments to 

a minimum.

• 

If  an  abnormal  endoscopic  image  appears  or  an  abnormal  function  occurs  but 

quickly  corrects  itself,  the  endoscope  may  have  malfunctioned.  In  this  case,  consider                             

abandoning the procedure because the irregularity can occur again, and the endoscope 

may not return to its normal condition. Stop the examination immediately and slowly 

withdraw the endoscope while viewing the endoscopic image. Otherwise, patient injury, 

bleeding, and/or perforation can result.

Over-insufflation & gas

• 

Confirm that the top hole of the insufflation/rinsing valve is not blocked. If the hole is 

blocked, air is fed continuously and patient pain, bleeding, and/or perforation can result. 

• 

Do not inflate air or a nonflammable gas excessively into the patient. This could cause gas 

embolism and/ or over-insufflation. 

6

Содержание 482001000

Страница 1: ...For use by trained clinicians physicians only For in hospital use For use with Ambu aBox Duodeno Instruction For Use Ambu aScope Duodeno...

Страница 2: ...b d e a c VAC CO2 b d a c CO2 VAC aBox aBox 1 9 13 8 7 2 4 3 6 5 12 11 10 Ambu is a registered trademark and aScope Duodeno is a trademark of Ambu A S 2...

Страница 3: ...2 Preparation Of The Endoscope 14 4 3 Attaching Ancillary Equipment To The Endoscope 14 4 4 Operating The aScope Duodeno 16 5 Ending A Procedure 17 5 1 Concluding A Patient Procedure 17 6 After Use 1...

Страница 4: ...doscopy and endoscopic treatment that are defined by the hospital s medical administrators or other official institutions such as academic societies on endoscopy follow those standards If there are no...

Страница 5: ...that fit properly and are long enough so that your skin is not exposed Please note that a new pair of gloves is required prior to each procedure Preparation for Use Before endoscopy remove any metall...

Страница 6: ...water level in the water bottle is too low replace the bottle with a new one An empty bottle can cause loss of rinsing and cooling function Always keep a new bottle of sterile water ready for this ca...

Страница 7: ...hold a piece of gauze around the accessory and the biopsy valve during withdrawal Do not let the endoscopic accessory hang down from the biopsy valve Doing so can create a space between the accessory...

Страница 8: ...instrument cannot be extended move the elevator control lever in the opposite direction of the Up direction to lower the elevator When the device is used with energized endoscopic devices leakage cur...

Страница 9: ...t The lens surface of the distal end is fragile and visual distortion may occur Do not coil the insertion tube or the umbilicus in a diameter of less than 12 cm Equipment damage can result Do not atte...

Страница 10: ...no is not available in all countries For detailed information please contact your local sales office 2 3 aScope Duodeno Description And Function The aScope Duodeno is a flexible endoscope with side vi...

Страница 11: ...move any fluids debris or gas from the patient 5 Insufflation rinsing valve The hole in this valve is covered to insufflate air and the valve is depressed to feed water for lens washing It also can be...

Страница 12: ...r Connects the endoscope to the suction tube of the suction pump 18 Insufflation connector Connects the endoscope to the insufflator 19 Lot Number Printed Lot Number of the endoscope 20 Endoscope stra...

Страница 13: ...d or damaged Device not for reuse Working length 124cm 48 8 of the aScope Duodeno Rx Only Prescription Device 4 Preparations For Use Numbers in gray circles below refer to illustrations on page 2 Befo...

Страница 14: ...that the sterile water bottle chosen should contain a minimum of 1000 ml in volume Please note that a new bottle of sterile water should be used at the start of each procedure Power up the aBox Duoden...

Страница 15: ...quipment back ON Connection to the Suction System Regardless of the vacuum source chosen the aScope Duodeno will require the source to provide a vacuum of min 7psi 50kPa and max 11psi 76kPa for the en...

Страница 16: ...e you insert the endoscope If necessary apply a medical grade lubricant to the insertion section Insert the distal end of the endoscope through the opening of the mouthpiece then from the mouth to the...

Страница 17: ...vance the endoscopic accessory slightly and move the elevator control lever in the Up direction Confirm that the accessory appears in the endoscopic image Manipulate the elevator control lever to adju...

Страница 18: ...for medical waste with electronic components This is a single use device so do not soak rinse or sterilize as it may leave harmful residues or cause malfunction of the device The design and material...

Страница 19: ...nstruments selected only on the basis of the working channel diameter will function in combination with the system For information regarding instruments compatibility please contact your Ambu sales re...

Страница 20: ...terile water connected or water bottle is empty Put a new bottle with sterile water in the bottle holder and connect it with the aScope Duodeno Rinsing tubing not properly connected Connect the rinsin...

Страница 21: ...new vacuum source suction pump Instrument channel blocked Flush sterile water with a syringe through the instru ment channel aScope Duodeno defective Withdraw the aScope Duodeno carefully and connect...

Страница 22: ...a or illumination failure Withdraw the aScope Duodeno and connect a new aScope Duodeno Cooling malfunction Connect a new bottle of sterile water and or insert the process water tubing so that it reach...

Страница 23: ...sted in the accompanying documentation are used The device is intended for use exclusively by trained medical personnel This device can cause radio interference or interference with the opera tion of...

Страница 24: ...idelines Electrostatic dischar ge ESD according to IEC 61000 4 2 8 kV contact discharge 15 kV air discharge 8 kV contact discharge 15 kV air discharge Floors should be wood or concrete or covered with...

Страница 25: ...3 V 3 V m Portable and mobile radio equipment should not be used closer to the unit including cords than the recom mended safety distance calculated according to the formula appropriate to the transmi...

Страница 26: ...cted directly to the public power grid that also supplies buildings used for residen tial purpose provided that the following warning is observed Warning This device is intended only for use by traine...

Страница 27: ...PM aScope Duodeno Rev 5 Manufacturer Ambu A S Baltorpbakken 13 DK 2750 Ballerup Denmark T 45 72 25 20 00 F 45 72 25 20 50 www ambu com...

Отзывы: