Air Techniques, Inc.
Page 6
AS12, AS22, AS50, AS70,
AS12C, AS22C, & AS50C
I n t e r n a t i o n a l
Service Clearance:
Allow 12" on all sides for all models .
Ambient Temperature:
Must not exceed 105°F
Air System Plumbing Connection:
3/8" F .N .P .T . Shut-off valve and a 6 ft . pressure
hose (supplied)
Air distribution piping for all models - 1/2", type "L"
or type "K" copper
If pipe volume is too great, more than 235 in
3
or
more than 100 ft . of 1/2" diameter pipe, a pressure
regulator should be installed between the main tank
and the distribution Set pressure to pressure switch
cut in value (factory set at 85 PSI) .
Environmental Operating:
Indoor use at altitudes up to 2000m . Temperature 5 to
40°C (41 to 105°F) .
Maximum relative humidity 80% for temperatures up to
31°C, decreasing linearly to 50% relative humidity at 40°C .
Supply voltage fluctuation of +/- 10% of nominal voltage .
Environmental Storage and Transport: Temperature, -18 to 65°C (0 to 150°F) .
Relative Humidity, 0 to 90% .
Classification - IEC 60601-1
Protection against electric shock (5 .1, 5 .2):
Class I
Applied Parts:
There are no Applied Parts .
Protection against harmful ingress of water (5 .3):
Ordinary, IPXO
Degree of safety in the presence of flammable
anesthetics mixture with air or with oxygen or with
nitrous oxide (5 .5):
Not suitable .
Mode of operation (5 .6):
50% Duty Cycle
SITE REQUIREMENTS