background image

Page 15

Air Techniques, Inc. 

AS12, AS22, AS50, AS70,

AS12C, AS22C, & AS50C 

I n t e r n a t i o n a l

MAINTENANCE

Important:

 

In dusty environments, the Intake Filter, PN 89938, may need to be changed more often than 

once a  year . 

 

Always dispose of the removed filter in accordance with local codes .

 

Filter Replacement - Yearly

 

Refer to Figure 5 for the location of filters to be replaced using the preventive maintenance 

kit for the specific 

AIRSTAR

 mode listed below .  Replace the filters and associated O-rings 

in accordance with the instructions provided with the kit .

Figure 5. 

AirStar Filter Location

Membrane Dryer

Top Membrane Filter 

P/N 87366 

Bottom Membrane Filter 

P/N 86787

Note:

  Top Cover O-ring, P/N 87368, is replaced 

with the Top Membrane Filter . 

See instructions provided with the kit .

Note:

  Bottom Cover O-ring, P/N 85529, is replaced 

with the Bottom Membrane Filter & Bowl . 

See instructions provided with the kit .

Compressor 

Air Intake Filters

P/N 89938R 

(1 each cylinder head)

Preventive Maintenance Kits Supplied Components

AirStar Model

AS12INT

AS12CINT

AS22INT

AS22CINT

AS50INT

AS50CINT AS70INT

85529

87366

87368

89938R - See Table 

for Kit Quantities

86787

Kit Part No

87351

87352

87353

87354

Component

Part No.

Qty

Qty

Qty

Qty

Compressor Air Intake Filter

89938R

1

2

4

6

Top Membrane Filter

87366

1

1

1

2

Bottom Membrane Filter & Bowl

86787

1

1

1

2

Top Cover O-ring

87368

1

1

1

2

Filter Bowl O-ring

85529

1

1

1

2

Содержание AirStar AS12CINT

Страница 1: ...Installation and Operation Manual DENTAL AIR SYSTEM AS12 AS22 AS50 AS70 AS12C AS22C AS50C International...

Страница 2: ...itude to cause electric shock MEDICAL ELECTRICAL EQUIPMENT WITH RESPECT TO ELECTRICAL SHOCK FIRE MECHANICAL AND OTHER SPECIFIED HAZARDS ONLY IN ACCORDANCE WITH UL 60601 1 CAN CSA C22 2 NO 601 1 66CA I...

Страница 3: ...odel and serial numbers available Then go to the Air Techniques web site www airtechniques com dental click the warranty registration link and complete the registration form This on line registration...

Страница 4: ...sembly Detail View Cooler Membrane Dryer Pressure Gauge Supplied Air Outlet Hose Tank Outlet Valve 5 Micron Filter Moisture Monitor Pressure Relief Valve Drain Valve Pressure Switch View A Front View...

Страница 5: ...at a time unless one head has failed and you are waiting for service The motor circuit breaker must be kept in the ON position and should not be used as a switch OPERATING INFORMATION Choosing the co...

Страница 6: ...n value factory set at 85 PSI Environmental Operating Indoor use at altitudes up to 2000m Temperature 5 to 40 C 41 to 105 F Maximum relative humidity 80 for temperatures up to 31 C decreasing linearly...

Страница 7: ...tubing for connection to the air intake of each compressor See Note 1 OUTSIDE AIR PIPE 2 Inch Pipe for Air Intake Must be protected from rain and animals Shroud Screen Screen Rear Clearance 12 inches...

Страница 8: ...l Site Requirements page 7 ELECTRICAL POWER CONNECTION AS12INT is supplied with an installed line cord with an IEC female connector that requires a 220V male IEC minimum 16AWG connector AS22INT AS50IN...

Страница 9: ...t and Remote Switch Panel When any 24V circuit exceeds 150 feet use 16 AWG 2 As shown 3 conductors of the 24V circuit cable from each compressor connect via the user supplied interconnect cable 3 The...

Страница 10: ...unning If not install the appropriate boost transformer and check that the correct main circuit breaker and wire size are being used Check pump up and recovery times as detailed below and compare to t...

Страница 11: ...Page 11 Air Techniques Inc AS12 AS22 AS50 AS70 AS12C AS22C AS50C Int e rna t iona l NOTES...

Страница 12: ...filter s not seated cor rectly a Remove filter s Replace if clogged or dirty When installing make sure filter chamber is clean and rubber flange on top of filter is pushed all the way down into the me...

Страница 13: ...until a click is heard Storage tank pressure should read 85 PSI on the pressure gauge Close the tank outlet valve turn on the power switch and verify the pump up time for your model AirStar Call your...

Страница 14: ...interrupted service Routine Inspection Monthly Clean exterior surfaces Check for abnormal noises and air leaks Make sure that no flammable corrosive or combustible materials are stored in the equipmen...

Страница 15: ...p Membrane Filter P N 87366 Bottom Membrane Filter P N 86787 Note Top Cover O ring P N 87368 is replaced with the Top Membrane Filter See instructions provided with the kit Note Bottom Cover O ring P...

Страница 16: ...REMOTE CONTROL PANEL w 24 V switches 1 Switch Plate Kit 2 Switch Plate Kit 3 Switch Plate Kit 4 Switch Plate Kit For all AirStars 53111 53251 53250 53133 SOUND COVER AS12INT AS22INT AS50INT AS70INT 85...

Страница 17: ...t Min 80 psi 2 5 5 0 10 0 15 0 Pump up Time 0 115 PSI 2 mins 55 secs 3 mins 10 secs 2 min 50 secs 2 min 40 secs Recovery Time 85 115 PSI 48 secs 47 secs 42 secs 40 secs Tank Size cu ft US Gal 0 8 6 1...

Страница 18: ...Air Techniques Inc Page 18 AS12 AS22 AS50 AS70 AS12C AS22C AS50C Inte rnational NOTES...

Страница 19: ...Page 19 Air Techniques Inc AS12 AS22 AS50 AS70 AS12C AS22C AS50C Int e rna t iona l NOTES...

Страница 20: ...tizer and Lotion Waterline Cleaner Evacuation System Cleaner Imaging Accessories Chemistry Processor Accessories For over 50 years Air Techniques has been a leading innovator and manufacturer of denta...

Отзывы: