Page 7
Air Techniques, Inc.
AS12, AS22, AS50, AS70,
AS12C, AS22C, & AS50C
I n t e r n a t i o n a l
Site Electrical Requirements
AS12INT &
AS12CINT
AS22INT &
AS22CINT
AS50INT &
AS50CINT
AS70INT
Nominal Supply Voltage (VAC, see note)
220
220
220
220
Frequency (Hz)
60
50/60
50/60
50/60
Full Load Amps
4
8
16
24
Minimum Panel Breaker Rating (Amps)
10
20
30
40
Minimum Wire Size (AWG)
16
12
10
8
Requirement
AirStar Model
Note:
Install a buck or boost transformer if actual facility service is above or below the
supply voltage fluctuation of ±10% of nominal voltage ratings listed .
SITE REQUIREMENTS
Figure 2. Overall Site Requirements
Remote 24 Volt Switch with
Pilot Light (sold separately)
3/8” FNPT End fitting
1/2” Copper Main Air Line
Control Cable 18 Gauge 4 Conductor
Connect to Pressure Switch .
See 24 Volt Circuit Connections .
13” max
Building Power
Supply Panel
Buck/Boost
Transformer
(optional)
Remote Air Intake Kit Manifold
- Kit
includes 2 inch PVC Pipe and flexible
hose with 70 inches of clear tubing for
connection to the air intake of each
compressor . See Note 1 .
OUTSIDE AIR PIPE
2-Inch Pipe for Air Intake .
Must be protected from rain and animals
Shroud &
Screen
Screen
Rear Clear
ance
12 inches minimum
from wall
Drip Leg with Drain Valve
See Note 2
36 Inch
Maximum
Height
*
Disconnect Needed
when Servicing .
Maximum 24
inches from
main air line
Membrane Dryer
Drain Valve
See Note 3
Notes:
1 .
Remote Air Intake Kit Manifold -
Refer to the
Optional Accessories listing, page 16, for the kit part number
corresponding to the
AirStar
model.
2 .
Remote Air Intake Drip Leg & Valve -
A drip leg
with drain valve must be installed at lower end of the
remote air intake pipe to collect condensation during
operation. Attach a drain tube to the drip leg valve to allow
drainage into floor drain/sink.
3 .
Membrane Dryer Drain Valve -
See Key Parts
Identification, page 4, and install the Drain Tube found in
either accessory kit P/N 87133 or P/N 87134 used to drain
moisture collected in the Membrane Dryer.
Side
Clearence
12 inch
es
Type
Style
AS12INT
IEC
Female
*
AS22INT
Hard
Wired
*
See Figure 3
for Connection
Details .
AS50INT
AS70INT
Use Country
Specific Hospital
Grade Receptacle
Only .