AIRSTAR

 generates 100% oil-less, ultra-dry dental air which protects valuable hand-

pieces from premature failure due to the effects of moist air and the build-up of oil residue . 

Because no oil is used for mechanical lubrication, there is no chance of introducing an oily film 

to a prepared surface which could compromise resin retention and restorations, wasting chair 

time . Most important, your patients’s health is protected with ultra-dry air that provides an envi-

ronment that is not conducive to bacterial growth .

The 

AIRSTAR

 utilizes a long stroke, small bore piston to compress the air . This piston is bonded 

with an anti-friction polymer to eliminate the need for oil . The air is forced through the Membrane 

Dryer System consisting of the cooler and the membrane . This system  removes moisture and 

air impurities providing the driest possible compressed air while maximizing performance . This 

100% ultra-dry air is reserved in the main storage tank for use by the operatory air system .  

The 

AIRSTAR

 features include:

• Virtually Maintenance Free

• Low Pressure Dew Point

• Maximum Dryness with Quadruple Filtered Air

• Uninterrupted Compressor Availability

• Compact size for space-saving installation

Since 1971, when Air Techniques pioneered the manufacture of oil-less air for dentistry, 

thousands  of dentists have depended on their 

AIRSTAR

 . Now that your practice has an 

AIRSTAR

 , you, too, can depend on the delivery of 100% oil-less, ultra-dry air and efficient, 

trouble-free operation . 

Quickly and easily register your new 

AIRSTAR

 on-line . Just have your product model and serial 

numbers available .  Then go to the Air Techniques web site, 

www.airtechniques.com/dental

click the 

warranty registration

 link and complete the registration form . This on-line registration 

ensures a record for the warranty period and helps us keep you informed of product updates 

and other valuable information . 

WARRANTY

CONGRATULATIONS

ON-LINE WARRANTY REGISTRATION

Each 

AIRSTAR

 is warranted to be free from defects in material and workmanship from the date 

of installation for a period of 2 years (24 months) on complete unit . 
All part and component returns and replacement equipment under warranty require a Return 

Materials Authorization (RMA) . Warranty returns must be received within three months of the 

RMA issue date . Items returned without an RMA, or included with other products for which an 

RMA has been issued, may be returned to the customer at the discretion of Air Techniques, lnc .
Any item returned under warranty, will be repaired or replaced at our option at no charge 

provided that our inspection shall indicate it to have been defective . Air Techniques, lnc . is 

not liable for indirect or consequential damages or loss of any nature in connection with this 

equipment . Dealer labor, shipping and handling charges are not covered by this warranty .
Warranty credit will not be applied to product returns that exhibit damage due to shipping, misuse, 

careless handling or repairs by unauthorized personnel . Credit, or partial credit, will not be issued 

until products/parts have been received and assessed . Warranty is void if product is installed or 

serviced by anyone other than an authorized Air Techniques' dealer or service personnel .
This warranty is in lieu of all other warranties expressed or implied . No representative or person 

is authorized to assume for us any liability in connection with the sale of our equipment .

Содержание AirStar AS12CINT

Страница 1: ...Installation and Operation Manual DENTAL AIR SYSTEM AS12 AS22 AS50 AS70 AS12C AS22C AS50C International...

Страница 2: ...itude to cause electric shock MEDICAL ELECTRICAL EQUIPMENT WITH RESPECT TO ELECTRICAL SHOCK FIRE MECHANICAL AND OTHER SPECIFIED HAZARDS ONLY IN ACCORDANCE WITH UL 60601 1 CAN CSA C22 2 NO 601 1 66CA I...

Страница 3: ...odel and serial numbers available Then go to the Air Techniques web site www airtechniques com dental click the warranty registration link and complete the registration form This on line registration...

Страница 4: ...sembly Detail View Cooler Membrane Dryer Pressure Gauge Supplied Air Outlet Hose Tank Outlet Valve 5 Micron Filter Moisture Monitor Pressure Relief Valve Drain Valve Pressure Switch View A Front View...

Страница 5: ...at a time unless one head has failed and you are waiting for service The motor circuit breaker must be kept in the ON position and should not be used as a switch OPERATING INFORMATION Choosing the co...

Страница 6: ...n value factory set at 85 PSI Environmental Operating Indoor use at altitudes up to 2000m Temperature 5 to 40 C 41 to 105 F Maximum relative humidity 80 for temperatures up to 31 C decreasing linearly...

Страница 7: ...tubing for connection to the air intake of each compressor See Note 1 OUTSIDE AIR PIPE 2 Inch Pipe for Air Intake Must be protected from rain and animals Shroud Screen Screen Rear Clearance 12 inches...

Страница 8: ...l Site Requirements page 7 ELECTRICAL POWER CONNECTION AS12INT is supplied with an installed line cord with an IEC female connector that requires a 220V male IEC minimum 16AWG connector AS22INT AS50IN...

Страница 9: ...t and Remote Switch Panel When any 24V circuit exceeds 150 feet use 16 AWG 2 As shown 3 conductors of the 24V circuit cable from each compressor connect via the user supplied interconnect cable 3 The...

Страница 10: ...unning If not install the appropriate boost transformer and check that the correct main circuit breaker and wire size are being used Check pump up and recovery times as detailed below and compare to t...

Страница 11: ...Page 11 Air Techniques Inc AS12 AS22 AS50 AS70 AS12C AS22C AS50C Int e rna t iona l NOTES...

Страница 12: ...filter s not seated cor rectly a Remove filter s Replace if clogged or dirty When installing make sure filter chamber is clean and rubber flange on top of filter is pushed all the way down into the me...

Страница 13: ...until a click is heard Storage tank pressure should read 85 PSI on the pressure gauge Close the tank outlet valve turn on the power switch and verify the pump up time for your model AirStar Call your...

Страница 14: ...interrupted service Routine Inspection Monthly Clean exterior surfaces Check for abnormal noises and air leaks Make sure that no flammable corrosive or combustible materials are stored in the equipmen...

Страница 15: ...p Membrane Filter P N 87366 Bottom Membrane Filter P N 86787 Note Top Cover O ring P N 87368 is replaced with the Top Membrane Filter See instructions provided with the kit Note Bottom Cover O ring P...

Страница 16: ...REMOTE CONTROL PANEL w 24 V switches 1 Switch Plate Kit 2 Switch Plate Kit 3 Switch Plate Kit 4 Switch Plate Kit For all AirStars 53111 53251 53250 53133 SOUND COVER AS12INT AS22INT AS50INT AS70INT 85...

Страница 17: ...t Min 80 psi 2 5 5 0 10 0 15 0 Pump up Time 0 115 PSI 2 mins 55 secs 3 mins 10 secs 2 min 50 secs 2 min 40 secs Recovery Time 85 115 PSI 48 secs 47 secs 42 secs 40 secs Tank Size cu ft US Gal 0 8 6 1...

Страница 18: ...Air Techniques Inc Page 18 AS12 AS22 AS50 AS70 AS12C AS22C AS50C Inte rnational NOTES...

Страница 19: ...Page 19 Air Techniques Inc AS12 AS22 AS50 AS70 AS12C AS22C AS50C Int e rna t iona l NOTES...

Страница 20: ...tizer and Lotion Waterline Cleaner Evacuation System Cleaner Imaging Accessories Chemistry Processor Accessories For over 50 years Air Techniques has been a leading innovator and manufacturer of denta...

Отзывы: