Page 13
Air Techniques, Inc.
AS12, AS22, AS50, AS70,
AS12C, AS22C, & AS50C
I n t e r n a t i o n a l
*
DIAGNOSTIC PROCEDURE FOR DEFECTIVE COMPRESSOR HEAD(S)
1. Put power switches in the OFF position.
2. Reset the circuit breaker if it was previously tripped.
3. Test heads by turning ONE on at a time. If the motor fails to start, or the circuit breaker trips, the problem may be in that compressor
head. Leave the power switch for the defective head in the OFF position. Call your Authorized Air Techniques dealer for service.
NOTE:
One head may be run TEMPORARILY while waiting for service.
4. If all heads run independently, but will not run together, check the line voltage. If the voltage is within the min./max. voltage required
in PRODUCT SPECIFICATIONS, call your Authorized Air Techniques dealer for service.
TROUBLESHOOTING
Problem
Possible Cause
Possible Solutions
5 . Compressor cycles
even when there is no
air demand from the
operatory .
a . Leak in the compressor .
a . Disconnect the main power supply .
Drain the storage tank slowly until a
“click” is heard . Storage tank pressure
should read 85 PSI on the pressure
gauge .
Close the tank outlet valve, turn on
the power switch and verify the pump-
up time for your model
AirStar
. Call
your authorized Air Techniques dealer
if the pump-uptime is incorrect . (See
Post Installation Check for pump-up
times .)
b . Leak in the office air system .
b . Look at the moisture monitor
(see KEY PARTS to locate) .
If it is blue, perform the following:
1 . With the
AirStar
’s power switch
ON, drain the storage tank to
85 PSI to start the compression
cycle .
2 . When the cycle shuts off at 115
PSI, close the storage tank outlet
valve .
3 . Wait 5 minutes and open the stor-
age tank outlet valve .
4 . If the pressure drops, the air leak
is in the office air system or deliv-
ery units and not in the
AirStar
.
Call your dealer or plumber for
service .
If it is pink, see #6 below
6 . Moisture monitor is not
blue (pink or white) .
a . Leak in the office air system .
a . If the moisture monitor is pink, there is
too much moisture in the system . Call
your authorized Air Techniques dealer
for service .
b . Compressor keeps cycling .
b . Check the SIZING GUIDE . There may
be excessive air demands placed on
the
AirStar
. A larger capacity model
may be required .