Installation Note N5230-90004
7
Step 1. Remove the Outer Cover
CAUTION
This procedure is best performed with the analyzer resting on its front handles
in the vertical position.
Do not place the analyzer on its front panel without the
handles
. This will damage the front panel assemblies.
Refer to
for this procedure.
1. Disconnect the power cord (if it has not already been disconnected).
2. With a T-20 TORX driver, remove the strap handles (item
①
) by loosening the screws
(item
②
) on both ends until the handle is free of the analyzer.
3. With a T-20 TORX driver, remove the four rear panel feet (item
③
) by removing the center
screws (item
④
).
4. Slide the four bottom feet (item
⑤
) off the cover.
5. Slide the cover off of the frame.
Figure 2
Outer Cover Removal
NOTE
The figure above shows the N5230A front panel and floppy disk drive. The
N5230C front panel has a slightly different appearance and does not
include a floppy disk drive.
Содержание N5230-60102
Страница 12: ...12 Installation Note N5230 90004 Figure 6 New Cable Installation...
Страница 18: ...18 Installation Note N5230 90004...