8
Installation Note N5230-90004
Step 2. Remove the Front Panel Assembly
Refer to
for this procedure.
1. With a T-10 TORX driver, remove the eight screws (item
①
) from the sides of the frame.
CAUTION
Before removing the front panel from the analyzer, lift and support the front of
the analyzer chassis.
2. Slide the front panel over the test port connectors.
3. Disconnect the front panel interface ribbon cable (item
②
). The front panel is now free
from the analyzer.
Figure 3
Front Panel Assembly Removal
NOTE
The figure above shows the N5230A front panel and floppy disk drive. The
N5230C front panel has a slightly different appearance and does not
include a floppy disk drive.