16
Installation Note N5230-90004
Step 9. Enable Options 014 and 1E1
Procedure Requirements
• The analyzer must be powered up and operating to perform this procedure.
• The Network Analyzer program must be running.
• A mouse must be connected to the analyzer for this procedure.
Option Enable Procedure
1. On the analyzer’s
System
menu, point to
Service
, and then click
Option Enable
.
2. In the
Select Desired Option
list, click
014 - Configurable Test Set
. Click
Enable
.
3. In the
Select Desired Option
list, click
1E1 - Source Attenuators
. Click
Enable
.
4. Click
Yes
in answer to the displayed question in the
Restart Analyzer?
box.
5. When the installation is complete, click
Exit
.
Option Verification Procedure
Once the analyzer has restarted and the Network Analyzer program is again running:
1. On the analyzer’s
Help
menu, click
About Network Analyzer
.
2. Verify that “014” and “1E1” are listed after “Options:” in the display. Click
OK
.
NOTE
If Options 014 and 1E1 have not been enabled, perform the
again. If the options are still not enabled, contact Agilent
Technologies. Refer to