Remove the inlet base plug.
a
Use a 7/16-inch open-end wrench (SS inlet) or 5/16-inch
and 1/4-inch wrenches (MM inlet) to loosen the sealing
screw at the inlet base. See
.
b
Gently depress the Guard chip heater assembly.
c
Remove the plug from the inlet base and discard.
Figure 12
Remove the inlet base plug
Remove plug
Multimode inlet
Split/Splitless inlet
Use two wrenches to
loosen inlet sealing
screw
Remove plug
Loosen inlet sealing
screw
Guard chip heater
assembly (shown
lowered)
Guard chip heater
assembly (shown
lowered)