Installing the GC
1
Agilent Intuvo 9000 GC Installation
35
you can scroll through the screens to select the consumables
data (source type, pump type, and so on).
Configure another MS type
If connected to an MS such as 5977A or 7000C/D, use the GC
touch screen to identify the connected MS’s IP address and to
select the part information (for example, source type and pump
type). Touch
Settings > Configuration > Detectors
, then select the
tab for the MSD. The MSD IP address entered here must exactly
match the IP address entered in the data system and at the MSD
keyboard.