112
Agilent Intuvo 9000 GC Installation
B
Cabling Diagrams and Remote Start/Stop
External event cable, G1530-60590
The external event cable has two passive relay contact closures
with two 24-volt control outputs. Devices connected to the
passive contact closures must be connected to their own power
sources.
The pin assignments for this cable are listed in
4
5
6
3
1
2
7 8
Wire terminations
Apply label
G1580-87200
Table 13
External events cable
Connector 1 pin
Signal name
Maximum rating
Connector 2, wire colorControlled by valve #
24 volts output
1
24 V output 1
150 mA
Yellow
5
2
24 V output 1
150 mA
Black
6
3
Ground
Red
4
Ground
White
Relay contact closures
(normally open)
5
Closure 1
48 V AC/DC, 250 mA
Orange
7
6
Closure 1
Green
7
Содержание Intuvo 9000
Страница 1: ...Agilent Technologies Agilent Intuvo 9000 Gas Chromatograph Installation and First Startup...
Страница 6: ...6 Agilent Intuvo 9000 GC Installation...
Страница 25: ...Installing the GC 1 Agilent Intuvo 9000 GC Installation 25 3 Reinstall the GC covers...
Страница 26: ...26 Agilent Intuvo 9000 GC Installation 1 Installing the GC Connect the power cord and LAN cable LAN...
Страница 27: ...Installing the GC 1 Agilent Intuvo 9000 GC Installation 27 Turn on the GC...
Страница 29: ...Installing the GC 1 Agilent Intuvo 9000 GC Installation 29...
Страница 52: ...52 Agilent Intuvo 9000 GC Installation 1 Installing the GC Detector cover Vent chimney Screw O Ring...
Страница 84: ...84 Agilent Intuvo 9000 GC Installation 1 Installing the GC 8 Close the GC front door...
Страница 114: ...Agilent Technologies...