Agilent Intuvo 9000 GC Installation
1
Installing the GC
REMOTE connector
Provides a port to remotely start and stop other instruments
using the APG protocol. A maximum of 10 instruments can be
synchronized using this connector. See
on page 100 for more detail.
EVENT connector
This connector provides two passive contact closures and two
24-volt outputs for controlling external devices.
BCD input connector
This connector provides two control relays and a BCD input for
a stream selection valve or a BCD generating device.
LAN connector
Standard Local Area Network (LAN) connector, for
communication with data systems and other devices via TCP/IP.
CAUTION
This connector is similar to the
EVENT
connector. Plugging a
non-BCD cable into the
BCD
connector can damage the GC.
Содержание Intuvo 9000
Страница 1: ...Agilent Technologies Agilent Intuvo 9000 Gas Chromatograph Installation and First Startup...
Страница 6: ...6 Agilent Intuvo 9000 GC Installation...
Страница 25: ...Installing the GC 1 Agilent Intuvo 9000 GC Installation 25 3 Reinstall the GC covers...
Страница 26: ...26 Agilent Intuvo 9000 GC Installation 1 Installing the GC Connect the power cord and LAN cable LAN...
Страница 27: ...Installing the GC 1 Agilent Intuvo 9000 GC Installation 27 Turn on the GC...
Страница 29: ...Installing the GC 1 Agilent Intuvo 9000 GC Installation 29...
Страница 52: ...52 Agilent Intuvo 9000 GC Installation 1 Installing the GC Detector cover Vent chimney Screw O Ring...
Страница 84: ...84 Agilent Intuvo 9000 GC Installation 1 Installing the GC 8 Close the GC front door...
Страница 114: ...Agilent Technologies...