:MEASure Commands
21
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
417
n/a
:MEASure:TVALue?
<value>,
[<slope>]<occurrence>
[,<source>] (see
page 469
)
<value> ::= voltage level that
the waveform must cross.
<slope> ::= direction of the
waveform when <value> is crossed.
<occurrence> ::= transitions
reported.
<source> ::= {CHANnel<n> |
FUNCtion<m> | MATH<m> |
WMEMory<r>} for DSO models
<source> ::= {CHANnel<n> |
DIGital<d> | FUNCtion<m> |
MATH<m> | WMEMory<r>} for MSO
models
<n> ::= 1 to (# analog channels)
in NR1 format
<m> ::= 1-4 in NR1 format
<r> ::= 1 to (# ref waveforms) in
NR1 format
<d> ::= 0 to (# digital channels
- 1) in NR1 format
<return_value> ::= time in
seconds of specified voltage
crossing in NR3 format
:MEASure:VAMPlitude
[<source>] (see
page 471
)
:MEASure:VAMPlitude?
[<source>] (see
page 471
)
<source> ::= {CHANnel<n> |
FUNCtion<m> | MATH<m> |
WMEMory<r>}
<n> ::= 1 to (# analog channels)
in NR1 format
<m> ::= 1-4 in NR1 format
<r> ::= 1 to (# ref waveforms) in
NR1 format
<return_value> ::= the amplitude
of the selected waveform in volts
in NR3 format
:MEASure:VAVerage
[<interval>][,<source
>] (see
page 472
)
:MEASure:VAVerage?
[<interval>][,<source
>] (see
page 472
)
<interval> ::= {CYCLe | DISPlay}
<source> ::= {CHANnel<n> |
FUNCtion<m> | MATH<m> |
WMEMory<r>}
<n> ::= 1 to (# analog channels)
in NR1 format
<m> ::= 1-4 in NR1 format
<r> ::= 1 to (# ref waveforms) in
NR1 format
<return_value> ::= calculated
average voltage in NR3 format
Table 96 :MEASure Commands Summary (continued)
Command
Query
Options and Query Returns
Содержание InfiniiVision 4000 X-Series
Страница 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Страница 32: ...32 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Страница 52: ...52 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 2 Setting Up ...
Страница 66: ...66 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 3 Getting Started ...
Страница 242: ...242 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 7 ACQuire Commands ...
Страница 252: ...252 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands ...
Страница 300: ...300 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 12 DEMO Commands ...
Страница 308: ...308 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 13 DIGital d Commands ...
Страница 330: ...330 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 15 DVM Commands ...
Страница 336: ...336 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 16 EXTernal Trigger Commands ...
Страница 368: ...368 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 17 FUNCtion m Commands ...
Страница 386: ...386 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 18 HARDcopy Commands ...
Страница 484: ...484 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 21 MEASure Commands ...
Страница 506: ...506 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 22 MEASure Power Commands ...
Страница 542: ...542 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 24 POD Commands POD1 DISP 0 THR 1 40E 00 ...
Страница 1098: ...1098 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 34 WGEN w Commands ...
Страница 1174: ...1174 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 37 Error Messages ...
Страница 1206: ...1206 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 39 Synchronizing Acquisitions ...
Страница 1336: ...1336 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 41 Programming Examples ...
Страница 1352: ...1352 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide Index ...