122
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
4
Commands Quick Reference
Table 29 :SBUS<n>:LIN Commands Summary
Command
Query
Options and Query Returns
:SBUS<n>:LIN:PARity
{{0 | OFF} | {1 |
ON}} (see
page 735
)
:SBUS<n>:LIN:PARity?
(see
page 735
)
{0 | 1}
:SBUS<n>:LIN:SAMPlepo
int <value> (see
page 736
)
:SBUS<n>:LIN:SAMPlepo
int? (see
page 736
)
<value> ::= {60 | 62.5 | 68 | 70
| 75 | 80 | 87.5} in NR3 format
:SBUS<n>:LIN:SIGNal:B
AUDrate <baudrate>
(see
page 737
)
:SBUS<n>:LIN:SIGNal:B
AUDrate? (see
page 737
)
<baudrate> ::= integer from 2400
to 625000 in 100 b/s increments
:SBUS<n>:LIN:SOURce
<source> (see
page 738
)
:SBUS<n>:LIN:SOURce?
(see
page 738
)
<source> ::= {CHANnel<n> |
EXTernal} for DSO models
<source> ::= {CHANnel<n> |
DIGital<d>} for MSO models
<n> ::= 1 to (# analog channels)
in NR1 format
<d> ::= 0 to (# digital channels
- 1) in NR1 format
:SBUS<n>:LIN:STANdard
<std> (see
page 739
)
:SBUS<n>:LIN:STANdard
? (see
page 739
)
<std> ::= {LIN13 | LIN20}
:SBUS<n>:LIN:SYNCbrea
k <value> (see
page 740
)
:SBUS<n>:LIN:SYNCbrea
k? (see
page 740
)
<value> ::= integer = {11 | 12 |
13}
:SBUS<n>:LIN:TRIGger
<condition> (see
page 741
)
:SBUS<n>:LIN:TRIGger?
(see
page 741
)
<condition> ::= {SYNCbreak | ID |
DATA}
:SBUS<n>:LIN:TRIGger:
ID <value> (see
page 742
)
:SBUS<n>:LIN:TRIGger:
ID? (see
page 742
)
<value> ::= 7-bit integer in
decimal, <nondecimal>, or
<string> from 0-63 or 0x00-0x3f
<nondecimal> ::= #Hnn where n ::=
{0,..,9 | A,..,F} for hexadecimal
<nondecimal> ::= #Bnn...n where n
::= {0 | 1} for binary
<string> ::= "0xnn" where n ::=
{0,..,9 | A,..,F} for hexadecimal
Содержание InfiniiVision 4000 X-Series
Страница 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Страница 32: ...32 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Страница 52: ...52 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 2 Setting Up ...
Страница 66: ...66 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 3 Getting Started ...
Страница 242: ...242 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 7 ACQuire Commands ...
Страница 252: ...252 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands ...
Страница 300: ...300 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 12 DEMO Commands ...
Страница 308: ...308 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 13 DIGital d Commands ...
Страница 330: ...330 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 15 DVM Commands ...
Страница 336: ...336 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 16 EXTernal Trigger Commands ...
Страница 368: ...368 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 17 FUNCtion m Commands ...
Страница 386: ...386 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 18 HARDcopy Commands ...
Страница 484: ...484 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 21 MEASure Commands ...
Страница 506: ...506 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 22 MEASure Power Commands ...
Страница 542: ...542 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 24 POD Commands POD1 DISP 0 THR 1 40E 00 ...
Страница 1098: ...1098 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 34 WGEN w Commands ...
Страница 1174: ...1174 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 37 Error Messages ...
Страница 1206: ...1206 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 39 Synchronizing Acquisitions ...
Страница 1336: ...1336 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 41 Programming Examples ...
Страница 1352: ...1352 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide Index ...