:TRIGger Commands
32
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
985
:TRIGger:PATTern:RANGe
(see
page 1208
)
Command Syntax
:TRIGger:PATTern:RANGe <less_than_time>[<suffix>],
<greater_than_time>[<suffix>]
<greater_than_time> ::= 10 ns to 9.99 seconds in NR3 format
<less_than_time> ::= 15 ns to 10 seconds in NR3 format
<suffix> ::= {s | ms | us | ns | ps}
The :TRIGger:PATTern:RANGe command sets the duration for the defined
pattern when the :TRIGger:PATTern:QUALifier command is set to INRange
or OUTRange. You can enter the parameters in any order — the smaller
value becomes the <greater_than_time> and the larger value becomes the
<less_than_time>.
Query Syntax
:TRIGger:PATTern:RANGe?
The :TRIGger:PATTern:RANGe? query returns the duration time for the
defined pattern.
Return Format
<less_than_time>,<greater_than_time><NL>
See Also
•
"Introduction to :TRIGger Commands"
on page 935
•
":TRIGger:PATTern"
on page 978
•
":TRIGger:PATTern:QUALifier"
on page 983
•
":TRIGger:MODE"
on page 944
Содержание InfiniiVision 4000 X-Series
Страница 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Страница 32: ...32 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Страница 52: ...52 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 2 Setting Up ...
Страница 66: ...66 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 3 Getting Started ...
Страница 242: ...242 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 7 ACQuire Commands ...
Страница 252: ...252 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands ...
Страница 300: ...300 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 12 DEMO Commands ...
Страница 308: ...308 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 13 DIGital d Commands ...
Страница 330: ...330 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 15 DVM Commands ...
Страница 336: ...336 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 16 EXTernal Trigger Commands ...
Страница 368: ...368 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 17 FUNCtion m Commands ...
Страница 386: ...386 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 18 HARDcopy Commands ...
Страница 484: ...484 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 21 MEASure Commands ...
Страница 506: ...506 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 22 MEASure Power Commands ...
Страница 542: ...542 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 24 POD Commands POD1 DISP 0 THR 1 40E 00 ...
Страница 1098: ...1098 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 34 WGEN w Commands ...
Страница 1174: ...1174 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 37 Error Messages ...
Страница 1206: ...1206 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 39 Synchronizing Acquisitions ...
Страница 1336: ...1336 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 41 Programming Examples ...
Страница 1352: ...1352 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide Index ...