:SBUS<n> Commands
28
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
745
:SBUS<n>:LIN:TRIGger:PATTern:DATA:LENGth
(see
page 1208
)
Command Syntax
:SBUS<n>:LIN:TRIGger:PATTern:DATA:LENGth <length>
<length> ::= integer from 1 to 8 in NR1 format
The :SBUS<n>:LIN:TRIGger:PATTern:DATA:LENGth command sets the
number of 8- bit bytes in the LIN data string. The number of bytes in the
string can be anywhere from 1 bytes to 8 bytes (64 bits). The value for
these bytes is set by the :SBUS<n>:LIN:TRIGger:PATTern:DATA command.
Query Syntax
:SBUS<n>:LIN:TRIGger:PATTern:DATA:LENGth?
The :SBUS<n>:LIN:TRIGger:PATTern:DATA:LENGth? query returns the
current LIN data pattern length setting.
Return Format
<count><NL>
<count> ::= integer from 1 to 8 in NR1 format
Errors
•
"- 241, Hardware missing"
on page 1169
See Also
•
"Introduction to :TRIGger Commands"
on page 935
•
":SBUS<n>:LIN:TRIGger:PATTern:DATA"
on page 743
•
":SBUS<n>:LIN:SOURce"
on page 738
Содержание InfiniiVision 4000 X-Series
Страница 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Страница 32: ...32 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Страница 52: ...52 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 2 Setting Up ...
Страница 66: ...66 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 3 Getting Started ...
Страница 242: ...242 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 7 ACQuire Commands ...
Страница 252: ...252 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands ...
Страница 300: ...300 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 12 DEMO Commands ...
Страница 308: ...308 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 13 DIGital d Commands ...
Страница 330: ...330 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 15 DVM Commands ...
Страница 336: ...336 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 16 EXTernal Trigger Commands ...
Страница 368: ...368 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 17 FUNCtion m Commands ...
Страница 386: ...386 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 18 HARDcopy Commands ...
Страница 484: ...484 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 21 MEASure Commands ...
Страница 506: ...506 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 22 MEASure Power Commands ...
Страница 542: ...542 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 24 POD Commands POD1 DISP 0 THR 1 40E 00 ...
Страница 1098: ...1098 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 34 WGEN w Commands ...
Страница 1174: ...1174 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 37 Error Messages ...
Страница 1206: ...1206 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 39 Synchronizing Acquisitions ...
Страница 1336: ...1336 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 41 Programming Examples ...
Страница 1352: ...1352 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide Index ...