:FUNCtion<m> Commands
17
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
357
:FUNCtion<m>:OFFSet
(see
page 1208
)
Command Syntax
:FUNCtion<m>:OFFSet <offset>
<m> ::= 1-4 in NR1 format
<offset> ::= the value at center screen in NR3 format.
The :FUNCtion<m>:OFFSet command sets the voltage or vertical value
represented at center screen for the selected function. The range of legal
values is gen/- 10 times the current scale of the selected function,
but will vary by function. If you set the offset to a value outside of the
legal range, the offset value is automatically set to the nearest legal value.
Query Syntax
:FUNCtion<m>:OFFSet?
The :FUNCtion<m>:OFFSet? query outputs the current offset value for the
selected function.
Return Format
<offset><NL>
<offset> ::= the value at center screen in NR3 format.
See Also
•
"Introduction to :FUNCtion<m> Commands"
on page 340
•
":FUNCtion<m>:RANGe"
on page 361
•
":FUNCtion<m>:REFerence"
on page 362
•
":FUNCtion<m>:SCALe"
on page 363
N O T E
The :FUNCtion<m>:OFFset command is equivalent to the :FUNCtion<m>:REFerence
command.
Содержание InfiniiVision 4000 X-Series
Страница 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Страница 32: ...32 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Страница 52: ...52 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 2 Setting Up ...
Страница 66: ...66 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 3 Getting Started ...
Страница 242: ...242 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 7 ACQuire Commands ...
Страница 252: ...252 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands ...
Страница 300: ...300 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 12 DEMO Commands ...
Страница 308: ...308 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 13 DIGital d Commands ...
Страница 330: ...330 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 15 DVM Commands ...
Страница 336: ...336 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 16 EXTernal Trigger Commands ...
Страница 368: ...368 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 17 FUNCtion m Commands ...
Страница 386: ...386 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 18 HARDcopy Commands ...
Страница 484: ...484 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 21 MEASure Commands ...
Страница 506: ...506 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 22 MEASure Power Commands ...
Страница 542: ...542 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 24 POD Commands POD1 DISP 0 THR 1 40E 00 ...
Страница 1098: ...1098 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 34 WGEN w Commands ...
Страница 1174: ...1174 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 37 Error Messages ...
Страница 1206: ...1206 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 39 Synchronizing Acquisitions ...
Страница 1336: ...1336 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 41 Programming Examples ...
Страница 1352: ...1352 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide Index ...