Chapter 2
55
Basic Operation
Configuring the RF Output
4.
Press
Sweep Time
to Auto.
The sweep time returns to its fastest allowable setting.
NOTE
When using an 8757D network analyzer in manual sweep mode, you must activate the signal
generator’s Manual Mode before using the
Manual Freq
softkey to control the sweep. Press
Sweep/List
>
More (2 of 3) > Manual Mode On.
Using Alternate Sweep
1.
Press the
Save
hardkey.
This opens the table editor and softkey menu for saving instrument states. Notice that the
Select
Reg
softkey is active. (For more information on saving instrument states refer to
“Using the
Instrument State Registers” on page 63
.)
2.
Turn the front panel knob until you find an
available
register and press
SAVE
. Remember this
saved register number. If no registers are available, you can write over an
in–use
register, by
pressing
Re–SAVE
.
NOTE
When you are using the PSG in a system with an 8757D network analyzer, you are limited to
using registers 1 through 9 in sequence 0 for saving and recalling states.
3.
Press
Sweep/List
>
Configure Ramp/Step Sweep
and enter new start and stop frequency values for the
ramp sweep.
4.
Press
Alternate Sweep Register
and turn the front panel knob to select the register number of the
previously saved sweep state.
5.
Press
Alternate Sweep Off On
to On.
The signal generator alternates between the original saved sweep and the current sweep. You may
need to adjust 8757D settings to effectively view both sweeps, such as setting channel 2 to
measure sensor A. Refer to
Figure 2- 7
.
Содержание E8257D
Страница 1: ...Agilent Technologies Agilent Technologies E8257D 67D E8663D PSG Signal Generators User s Guide ...
Страница 12: ...Contents xii ...
Страница 50: ...36 Chapter 1 Signal Generator Overview Rear Panel ...
Страница 92: ...78 Chapter 2 Basic Operation Using the Web Server ...
Страница 136: ...122 Chapter 3 Basic Digital Operation Using Waveform Clipping Figure 3 15 Rectangular Clipping ...
Страница 164: ...150 Chapter 4 Optimizing Performance Optimizing Phase Noise and Harmonics Below 3 2 GHz Option UNX UNY ...
Страница 194: ...180 Chapter 6 Custom Arb Waveform Generator Configuring Hardware ...
Страница 234: ...220 Chapter 8 GPS Modulation Option 409 Real Time GPS ...
Страница 244: ...230 Chapter 9 Multitone Waveform Generator Creating Viewing and Optimizing Multitone Waveforms ...
Страница 288: ...274 Chapter 12 Peripheral Devices Millimeter Wave Source Modules Figure 12 21 E8257D PSG without Option 1EA 1EU or 521 ...