Chapter 12
Peripheral Devices
N5102A Digital Signal Interface Module
This error is reported when the output FIFO is overflowing in the
digital module. This error can be generated if an external clock
or its reference is not set up properly, or if the internal VCO is
unlocked.
806
Digital module output FIFO underflow error; There are not
enough samples being produced for the current clock rate.
Verify that the digital module clock is set up properly.
This error is reported when the output FIFO is underflowing in the digital
module. This error can be generated if an external clock or its reference is not
set up properly, or if the internal VCO is unlocked.
2.
If the port configuration is parallel or parallel interleaved, using an IQ signal type, press the
Clocks Per Sample
softkey.
Notice that multiple clocks per sample can be selected. Some DACs require the ability to clock
multiple times for each sample; having a clocks per sample value greater than one reduces the
sample rate by a factor equal to the selected number of clocks per sample. The sample rate is
viewed on the first- level and Data Setup softkey menus.
3.
Select the clocks per sample value to fit the test.
4.
Press the
Clock Source
softkey.
From this menu, select the clock signal source. With each selection, the clock routing display in
the signal generator clock setup menu will change to reflect the current clock source. This will be
indicated by a change in the graphic.
5.
Select the clock source.
If External or Device is Selected
Press the
Clock Rate
softkey and enter the clock rate of the externally applied clock signal.
NOTE
The clock phase and clock skew may need to be adjusted each time the clock rate setting is
changed. Refer to
“Clock Timing for Phase and Skew Adjustments” on page 251
.
For the
External
selection, the signal is supplied by an external clock source and applied to the Ext
Clock In connector. For the
Device
selection, the clock signal is supplied through the Device
Interface connector, generally by the device under test.
If Internal is Selected
Using an external frequency reference, the N5102A module generates its own internal clock signal.
The reference frequency signal must be applied to the Freq Ref connector on the digital module.
a.
Press the
Reference Frequency
softkey and enter the frequency of the externally applied frequency
reference.
b.
Press the
Clock Rate
softkey and enter the appropriate clock rate.
Table 12- 8
provides a quick view of the settings and connections associated with each clock
source selection.
Содержание E8257D
Страница 1: ...Agilent Technologies Agilent Technologies E8257D 67D E8663D PSG Signal Generators User s Guide ...
Страница 12: ...Contents xii ...
Страница 50: ...36 Chapter 1 Signal Generator Overview Rear Panel ...
Страница 92: ...78 Chapter 2 Basic Operation Using the Web Server ...
Страница 136: ...122 Chapter 3 Basic Digital Operation Using Waveform Clipping Figure 3 15 Rectangular Clipping ...
Страница 164: ...150 Chapter 4 Optimizing Performance Optimizing Phase Noise and Harmonics Below 3 2 GHz Option UNX UNY ...
Страница 194: ...180 Chapter 6 Custom Arb Waveform Generator Configuring Hardware ...
Страница 234: ...220 Chapter 8 GPS Modulation Option 409 Real Time GPS ...
Страница 244: ...230 Chapter 9 Multitone Waveform Generator Creating Viewing and Optimizing Multitone Waveforms ...
Страница 288: ...274 Chapter 12 Peripheral Devices Millimeter Wave Source Modules Figure 12 21 E8257D PSG without Option 1EA 1EU or 521 ...