![Agilent Technologies E8257D Скачать руководство пользователя страница 56](http://html.mh-extra.com/html/agilent-technologies/e8257d/e8257d_user-manual_2867504056.webp)
42
Chapter 2
Basic Operation
Configuring the RF Output
Configuring the RF Output
This section provides information on how to create continuous wave and swept RF (
page 45
) outputs.
It also has information on using a mm–Wave source module to extend the signal generator’s
frequency range (
page 59
).
Configuring a Continuous Wave RF Output
These procedures demonstrate how to set the following parameters:
•
setting the RF output frequency (
page 42
)
•
setting the frequency reference and frequency offset (
page 43
)
•
setting the low pass filter (Options 1EH and 521) (
page 44
)
•
setting the RF output amplitude (
page 44
)
•
setting the amplitude reference and amplitude offset (
page 44
)
•
setting the optimize signal to noise ratio (
page 45
)
Setting the RF Output Frequency
Set the RF output frequency to 700 MHz, and increment or decrement the output frequency in 1 MHz
steps.
1.
Return the signal generator to the factory–defined state: Press
Preset
.
NOTE
You can change the preset condition of the signal generator to a user–defined state. For
these examples, however, use the factory–defined preset state (set the
Preset Normal User
softkey
in the Utility menu to Normal).
2.
Observe the
FREQUENCY
area of the display (in the upper left–hand corner).
The value displayed is the maximum specified frequency of the signal generator.
3.
Press
RF On/Off
.
The
RF On/Off
hardkey must be pressed before the RF signal is available at the
RF OUTPUT connector. The display annunciator changes from
RF OFF
to
RF ON
. The maximum
specified frequency should be output at the RF OUTPUT connector (at the signal generator’s
minimum power level).
4.
Press
Frequency
>
700
>
MHz
.
The 700 MHz RF frequency should be displayed in the
FREQUENCY
area of the display and also in
the active entry area.
5.
Press
Frequency
>
Incr Set
>
1
>
MHz
.
This changes the frequency increment value to 1 MHz.
6.
Press the up arrow key.
Each press of the up arrow key increases the frequency by the increment value last set with the
Incr Set
hardkey. The increment value is displayed in the active entry area.
Содержание E8257D
Страница 1: ...Agilent Technologies Agilent Technologies E8257D 67D E8663D PSG Signal Generators User s Guide ...
Страница 12: ...Contents xii ...
Страница 50: ...36 Chapter 1 Signal Generator Overview Rear Panel ...
Страница 92: ...78 Chapter 2 Basic Operation Using the Web Server ...
Страница 136: ...122 Chapter 3 Basic Digital Operation Using Waveform Clipping Figure 3 15 Rectangular Clipping ...
Страница 164: ...150 Chapter 4 Optimizing Performance Optimizing Phase Noise and Harmonics Below 3 2 GHz Option UNX UNY ...
Страница 194: ...180 Chapter 6 Custom Arb Waveform Generator Configuring Hardware ...
Страница 234: ...220 Chapter 8 GPS Modulation Option 409 Real Time GPS ...
Страница 244: ...230 Chapter 9 Multitone Waveform Generator Creating Viewing and Optimizing Multitone Waveforms ...
Страница 288: ...274 Chapter 12 Peripheral Devices Millimeter Wave Source Modules Figure 12 21 E8257D PSG without Option 1EA 1EU or 521 ...