146
Chapter 4
Optimizing Performance
Using the Option 521 Detector Calibration (Option 521)
Recalling and Applying a User Flatness Correction Array
Before performing the steps in this section, complete the section
“Creating a User Flatness Correction
Array with a mm–Wave Source Module” on page 140
.
1.
Press
Preset
.
2.
Press
Amplitude
>
More (1 of 2)
>
User Flatness
>
Configure Cal Array
>
More (1 of 2)
>
Preset List
>
Confirm Preset
.
3.
Press
More (2 of 2)
>
Load/Store
.
4.
Ensure that the file
FLATCAL2
is highlighted.
5.
Press
Load From Selected File
>
Confirm Load From File
.
This populates the user flatness correction array with the data contained in the file
FLATCAL2
.
The user flatness correction array title displays
User Flatness: FLATCAL2
.
6.
Press
Return
>
Flatness Off On
.
This activates flatness correction using the data contained in the file
FLATCAL2
.
Using the Option 521 Detector Calibration (Option 521)
CAUTION
Using the Option 521 Detector Calibration softkey changes the factory calibration. Avoid
using this feature unless following a service center’s recommendations.
This calibration should be done whenever the instrument temperature changes beyond
specified parameters or and when greater amplitude flatness is required around the
500 MHz frequency transition. Refer to the
Data Sheet
.
NOTE
If the Option 521 Detector Calibration is inadvertently used, the PSG has the ability to recall
the factory amplitude flatness calibration. Refer to
“Restoring the Factory Flatness
Calibration”
.
The purpose of this calibration is to improve the 500 MHz frequency transition to compensate for
changes in temperature. The 500 MHz frequency transition can degrade as the temperature changes.
By using the detector calibration, when the instrument temperature change occurs, the PSG’s flatness
performance can be optimized. Refer to the
Data Sheet
.
Running the Option 521 Detector Calibration
Following the following steps to run the Detector calibration:
•
Press
Amplitude > More > More > Option 521 Detector Calibration > Execute
Restoring the Factory Flatness Calibration
The instrument’s original factory calibration settings can be restored by the following steps:
•
Press
Utility > Power On/Preset > Restore System Defaults
Содержание E8257D
Страница 1: ...Agilent Technologies Agilent Technologies E8257D 67D E8663D PSG Signal Generators User s Guide ...
Страница 12: ...Contents xii ...
Страница 50: ...36 Chapter 1 Signal Generator Overview Rear Panel ...
Страница 92: ...78 Chapter 2 Basic Operation Using the Web Server ...
Страница 136: ...122 Chapter 3 Basic Digital Operation Using Waveform Clipping Figure 3 15 Rectangular Clipping ...
Страница 164: ...150 Chapter 4 Optimizing Performance Optimizing Phase Noise and Harmonics Below 3 2 GHz Option UNX UNY ...
Страница 194: ...180 Chapter 6 Custom Arb Waveform Generator Configuring Hardware ...
Страница 234: ...220 Chapter 8 GPS Modulation Option 409 Real Time GPS ...
Страница 244: ...230 Chapter 9 Multitone Waveform Generator Creating Viewing and Optimizing Multitone Waveforms ...
Страница 288: ...274 Chapter 12 Peripheral Devices Millimeter Wave Source Modules Figure 12 21 E8257D PSG without Option 1EA 1EU or 521 ...