Chapter 3
121
Basic Digital Operation
Using Waveform Clipping
appears as a rectangle in the vector representation. With either method, the objective is to clip the
waveform to a level that effectively reduces spectral regrowth, but does
not
compromise the integrity
of the signal.
Figure 3- 16 on page 123
uses two complementary cumulative distribution plots to show
the reduction in peak–to–average power that occurs after applying circular clipping to a waveform.
The lower you set the clipping value, the lower the peak power that is passed (or the more the signal
is clipped). Often, the peaks can be clipped successfully without substantially interfering with the rest
of the waveform. Data that might be lost in the clipping process is salvaged because of the error
correction inherent in the coded systems. If you clip too much of the waveform, however, lost data is
irrecoverable. You may have to try several clipping settings to find a percentage that works well.
Figure 3-14
Circular Clipping
Содержание E8257D
Страница 1: ...Agilent Technologies Agilent Technologies E8257D 67D E8663D PSG Signal Generators User s Guide ...
Страница 12: ...Contents xii ...
Страница 50: ...36 Chapter 1 Signal Generator Overview Rear Panel ...
Страница 92: ...78 Chapter 2 Basic Operation Using the Web Server ...
Страница 136: ...122 Chapter 3 Basic Digital Operation Using Waveform Clipping Figure 3 15 Rectangular Clipping ...
Страница 164: ...150 Chapter 4 Optimizing Performance Optimizing Phase Noise and Harmonics Below 3 2 GHz Option UNX UNY ...
Страница 194: ...180 Chapter 6 Custom Arb Waveform Generator Configuring Hardware ...
Страница 234: ...220 Chapter 8 GPS Modulation Option 409 Real Time GPS ...
Страница 244: ...230 Chapter 9 Multitone Waveform Generator Creating Viewing and Optimizing Multitone Waveforms ...
Страница 288: ...274 Chapter 12 Peripheral Devices Millimeter Wave Source Modules Figure 12 21 E8257D PSG without Option 1EA 1EU or 521 ...