![Agilent Technologies E8257D Скачать руководство пользователя страница 159](http://html.mh-extra.com/html/agilent-technologies/e8257d/e8257d_user-manual_2867504159.webp)
Chapter 4
145
Optimizing Performance
Creating and Applying User Flatness Correction
and their calculated amplitude correction values. The user flatness correction array title displays
User Flatness: (UNSTORED)
indicating that the current user flatness correction array data has
not been saved to the memory catalog.
Performing the User Flatness Correction Manually
If you are not using an Agilent E4416A/17A/18B/19B power meter, or if your power meter does not
have a GPIB interface, complete the steps in this section and then continue with the user flatness
correction tutorial.
1.
Press
More (1 of 2)
>
User Flatness
>
Configure Cal Array
.
This opens the User Flatness table editor and places the cursor over the frequency value
(26.5 GHz) for row 1. The RF output changes to the frequency value of the table row containing
the cursor and
26.500 000 000 00
is displayed in the
AMPLITUDE
area of the display.
2.
Observe and record the measured value from the power meter.
3.
Subtract the measured value from 0 dBm.
4.
Move the table cursor over the correction value in row 1.
5.
Press
Edit Item
> enter the difference value from step 3 >
dB
.
The signal generator adjusts the RF output amplitude based on the correction value entered.
6.
Repeat steps 2 through 5 until the power meter reads 0 dBm.
7.
Use the down arrow key to place the cursor over the frequency value for the next row. The RF
output changes to the frequency value highlighted by the cursor, as shown in the
AMPLITUDE
area
of the display.
8.
Repeat steps 2 through 7 for each entry in the User Flatness table.
Save the User Flatness Correction Data to the Memory Catalog
This process allows you to save the user flatness correction data as a file in the signal generator’s
memory catalog. With several user flatness correction files saved to the memory catalog, specific files
can be recalled, loaded into the correction array, and applied to the RF output to satisfy various RF
output flatness requirements.
1.
Press
Load/Store
.
2.
Press
Store to File
.
3.
Enter the file name
FLATCAL2
using the alphanumeric softkeys and the numeric keypad.
4.
Press
Enter
.
The user flatness correction array file FLATCAL2 is now stored in the memory catalog as a
UFLT
file.
Applying the User Flatness Correction Array
1.
Press
Return
>
Return
>
Flatness Off On
.
This applies the user flatness correction array to the RF output. The
UF
indicator is activated in
the
AMPLITUDE
section of the signal generator’s display and the frequency correction data
contained in the correction array is applied to the RF output amplitude of the mm–wave source
module.
Содержание E8257D
Страница 1: ...Agilent Technologies Agilent Technologies E8257D 67D E8663D PSG Signal Generators User s Guide ...
Страница 12: ...Contents xii ...
Страница 50: ...36 Chapter 1 Signal Generator Overview Rear Panel ...
Страница 92: ...78 Chapter 2 Basic Operation Using the Web Server ...
Страница 136: ...122 Chapter 3 Basic Digital Operation Using Waveform Clipping Figure 3 15 Rectangular Clipping ...
Страница 164: ...150 Chapter 4 Optimizing Performance Optimizing Phase Noise and Harmonics Below 3 2 GHz Option UNX UNY ...
Страница 194: ...180 Chapter 6 Custom Arb Waveform Generator Configuring Hardware ...
Страница 234: ...220 Chapter 8 GPS Modulation Option 409 Real Time GPS ...
Страница 244: ...230 Chapter 9 Multitone Waveform Generator Creating Viewing and Optimizing Multitone Waveforms ...
Страница 288: ...274 Chapter 12 Peripheral Devices Millimeter Wave Source Modules Figure 12 21 E8257D PSG without Option 1EA 1EU or 521 ...