246
Chapter 12
Peripheral Devices
N5102A Digital Signal Interface Module
Clock Timing for Parallel Data
Some components require multiple clocks during a single sample period. (A sample period consists of
an I and Q sample). For parallel data transmissions, you can select one, two, or four clocks per
sample. For clocks per sample greater than one, the I and Q samples are held constant to
accommodate the additional clock periods. This reduces the sample rate relative to the clock rate by
a factor equal to the clocks per sample selection. For example, when four is selected, the sample rate
is reduced by a factor of four (sample rate to clock rate ratio).
Figure 12- 4
demonstrates the clock
timing for each clocks per sample selection. For input mode, the clocks per sample setting is always
one.
Externally Supplied Clock
NOTE
: Use only one of the two signal generator frequency reference inputs.
Содержание E8257D
Страница 1: ...Agilent Technologies Agilent Technologies E8257D 67D E8663D PSG Signal Generators User s Guide ...
Страница 12: ...Contents xii ...
Страница 50: ...36 Chapter 1 Signal Generator Overview Rear Panel ...
Страница 92: ...78 Chapter 2 Basic Operation Using the Web Server ...
Страница 136: ...122 Chapter 3 Basic Digital Operation Using Waveform Clipping Figure 3 15 Rectangular Clipping ...
Страница 164: ...150 Chapter 4 Optimizing Performance Optimizing Phase Noise and Harmonics Below 3 2 GHz Option UNX UNY ...
Страница 194: ...180 Chapter 6 Custom Arb Waveform Generator Configuring Hardware ...
Страница 234: ...220 Chapter 8 GPS Modulation Option 409 Real Time GPS ...
Страница 244: ...230 Chapter 9 Multitone Waveform Generator Creating Viewing and Optimizing Multitone Waveforms ...
Страница 288: ...274 Chapter 12 Peripheral Devices Millimeter Wave Source Modules Figure 12 21 E8257D PSG without Option 1EA 1EU or 521 ...