Switching Channels
For general purpose relay operation, you can connect or disconnect a load
by opening or closing specified channel relays. By adding jumpers and/or
pullup/pulldown resistors, the switch can be configured for relay control or
digital output operations.
Use CLOS <channel_list> to connect a channel’s normally open (NO)
terminal to its common (C) terminal or use OPEN <channel_list> to connect
a channel’s normally closed (NC) contact to its common (C) terminal.
<channel_list> has the form (@ccnn) where cc = card number (01-99) and
nn = channel number (00-15).
Example: Voltage
Switching
This example closes channel 00 of a Form C Switch to switch the load
voltage (E) from Load 1 to Load 2. When the channel relay is open, the load
voltage is applied to Load 1. When the relay is closed, the voltage is applied
to Load 2. See Figure 3-1 for typical user connections. To close channel 00,
execute:
CLOS (@100)
Close channel 00 relay (connect
NO to C). 1 is the card number and
00 is the channel number.
To open channel 00, use
OPEN (@100)
.
Figure 3-1. Example: Voltage Switching
26 Using the Agilent E1364A Form C Switch
Chapter 3
Содержание E1364A
Страница 2: ......
Страница 6: ...Notes 4 Agilent E1364A Switch Module Table of Contents ...
Страница 10: ...Notes 8 Agilent E1364A 16 Channel Form C Switch User s Manual ...
Страница 11: ...Notes Agilent E1364A 16 Channel Form C Switch User s Manual 9 ...
Страница 12: ...Notes 10 Agilent E1364A 16 Channel Form C Switch User s Manual ...
Страница 18: ...16 Getting Started with the Agilent E1364A Chapter 1 ...
Страница 21: ...Figure 2 2 Wiring the Form C Switch Terminal Block Chapter 2 Configuring the Agilent E1364A Form C Switch 19 ...
Страница 70: ...68 Agilent E1364A Form C Switch Registers Appendix B ...
Страница 72: ...70 Agilent E1364A Form C Switch Error Messages Appendix C ...
Страница 76: ...74 Agilent E1364A 16 Channel Form C Switch Index ...