The ARM subsystem selects the number of scanning cycles (1 to 32767) for
each INITiate command.
Subsystem Syntax
ARM
:COUNt
<number> MIN | MAX
:COUNt?
[MIN
|
MAX]
:COUNt
ARM:COUNt <number> MIN | MAX
allows scanning cycles to occur a
multiple of times (1 to 32767) with one INITiate command when
INITiate:CONTinuous OFF | 0 is set. MIN sets 1 cycle, MAX sets 32767
cycles.
Parameters
Parameter Name
Parameter Type
Range of Values
<number>
numeric
1-32767 l MIN l MAX
Comments
•
Number of Scans: Use only values between 1 and 32767 for the
number of scanning cycles.
•
Related Commands: ABORt, INITiate[:IMMediate]
•
*RST Condition: ARM:COUNt 1
Example
Setting Ten Scanning Cycles:
This example sets a Form C Switch for 10 scans of channels 00 through 03.
When the scan sequence completes, channels 00 through 03 (relays 00
through 03) are closed.
ARM COUN 10
Set 10 scans per INIT command
SCAN (@100:103)
Scan channels 00-03
INIT
Start scan, close channel 00
Chapter 5
Agilent E1364A Form C Switch Command Reference 45
Содержание E1364A
Страница 2: ......
Страница 6: ...Notes 4 Agilent E1364A Switch Module Table of Contents ...
Страница 10: ...Notes 8 Agilent E1364A 16 Channel Form C Switch User s Manual ...
Страница 11: ...Notes Agilent E1364A 16 Channel Form C Switch User s Manual 9 ...
Страница 12: ...Notes 10 Agilent E1364A 16 Channel Form C Switch User s Manual ...
Страница 18: ...16 Getting Started with the Agilent E1364A Chapter 1 ...
Страница 21: ...Figure 2 2 Wiring the Form C Switch Terminal Block Chapter 2 Configuring the Agilent E1364A Form C Switch 19 ...
Страница 70: ...68 Agilent E1364A Form C Switch Registers Appendix B ...
Страница 72: ...70 Agilent E1364A Form C Switch Error Messages Appendix C ...
Страница 76: ...74 Agilent E1364A 16 Channel Form C Switch Index ...