Typical Switch Configurations
Typical Form C Switch configurations are:
•
General purpose relay
•
Relay control
•
Digital output
General Purpose
Relay Configuration
As factory configured, the Form C Switch is set for general purpose relay
configuration. For this configuration, you can switch channels by opening or
closing channel relays or you can scan a set of channels.
Example: General
Purpose Relay
Configuration
Figure 2-3 shows a typical general purpose relay configuration for channel
00. When the relay is open, the NC terminal is connected to the C terminal
and Load 1 is connected. When the relay is closed, the NO terminal is
connected to the C terminal and Load 2 is connected.
Relay Control
Configuration
By connecting jumpers, you can use the Form C Switch to drive (control)
external relays. The source can be +5V/NO, +5V/NC, +12V/NO, or
+12V/NC depending on the jumpers connected on the switch. (See Figure
1-1 in Chapter 1 for jumper numbers and see Figure 2-6 for jumper
locations.)
Note
When connecting the +5V or +12V backplane sources, the total current
draw should not exceed 1 A per Form C Switch for the +5V supply or 0.5 A
per Form C Switch for the +12V supply. You should fuse all external
equipment to ensure excess current is not drawn.
Figure 2-3. Example: General Purpose Relay Configuration
20 Configuring the Agilent E1364A Form C Switch
Chapter 2
Содержание E1364A
Страница 2: ......
Страница 6: ...Notes 4 Agilent E1364A Switch Module Table of Contents ...
Страница 10: ...Notes 8 Agilent E1364A 16 Channel Form C Switch User s Manual ...
Страница 11: ...Notes Agilent E1364A 16 Channel Form C Switch User s Manual 9 ...
Страница 12: ...Notes 10 Agilent E1364A 16 Channel Form C Switch User s Manual ...
Страница 18: ...16 Getting Started with the Agilent E1364A Chapter 1 ...
Страница 21: ...Figure 2 2 Wiring the Form C Switch Terminal Block Chapter 2 Configuring the Agilent E1364A Form C Switch 19 ...
Страница 70: ...68 Agilent E1364A Form C Switch Registers Appendix B ...
Страница 72: ...70 Agilent E1364A Form C Switch Error Messages Appendix C ...
Страница 76: ...74 Agilent E1364A 16 Channel Form C Switch Index ...