Digital Output
Configuration
By connecting jumpers and installing pullup/pulldown resistors, you can
configure the switch as a digital output device.
Note
When connecting the + 5V or + 12V backplane sources for digital output
operation, you must install pullup resistors on the Form C Switch for the
supplies used. Also, the total current drawn by user circuits should not
exceed 1 A per Form C Switch for the + 5V supply or 0.5 A per Form C
Switch for the + 12V supply. You should fuse all external equipment to
ensure excess current is not drawn.
Example: Digital
Output Configuration
Figure 2-5 shows channel 00 configured for digital output operation. When
the channel 00 relay is open (NC connected to C), point A is at + 5V. When
the channel 00 relay is closed (NO connected to C), point A is at 0V.
Figure 2-5. Example: Digital Output Configuration
22 Configuring the Agilent E1364A Form C Switch
Chapter 2
Содержание E1364A
Страница 2: ......
Страница 6: ...Notes 4 Agilent E1364A Switch Module Table of Contents ...
Страница 10: ...Notes 8 Agilent E1364A 16 Channel Form C Switch User s Manual ...
Страница 11: ...Notes Agilent E1364A 16 Channel Form C Switch User s Manual 9 ...
Страница 12: ...Notes 10 Agilent E1364A 16 Channel Form C Switch User s Manual ...
Страница 18: ...16 Getting Started with the Agilent E1364A Chapter 1 ...
Страница 21: ...Figure 2 2 Wiring the Form C Switch Terminal Block Chapter 2 Configuring the Agilent E1364A Form C Switch 19 ...
Страница 70: ...68 Agilent E1364A Form C Switch Registers Appendix B ...
Страница 72: ...70 Agilent E1364A Form C Switch Error Messages Appendix C ...
Страница 76: ...74 Agilent E1364A 16 Channel Form C Switch Index ...