OPEN?
[ROUTe:]OPEN? <channel_list>
returns the current state of the channel(s)
queried. channel_list has the form (@ccnn) (see [ROUTe:]OPEN for
definition). The command returns 1 if channel(s) are open or returns 0 if
channel(s) are closed.
Comments
•
Query is Software Readback: The ROUTe:OPEN? command
returns the current software state of the channels specified. It does
not account for relay hardware failures. A maximum of 127 channels
at a time can be queried for a multi-module switchbox.
Example
Query Channel Open State:
This example opens channels 100 and 213 of a two-module switchbox and
queries channel 213 state. Since channel 213 is programmed to be open, 1 is
returned.
OPEN (@100,213)
Open channels 100 end 213
OPEN? (@213)
Query channel 213 state
SCAN
[ROUTe:]SCAN <channel_list>
defines the channels to be scanned.
channel_list has the form (@ccnn) where cc = card number (01-99) and nn
= channel number (00-15).
Parameters
Parameter Name
Parameter Type
Range of Values
<channel_list>
numeric
cc00 - cc15
Comments
•
Defining Scan List: When [ROUTe:]SCAN is executed, the channel
list is checked for valid card and channel numbers. An error is
generated for an invalid channel list.
•
Scanning Channels: You can scan single channels (@ccnn);
multiple channels (@ccnn,ccnn,...); sequential channels
(@ccnn:ccnn); groups of sequential channels
(@ccnn:ccnn,ccnn:ccnn); or any combination.
•
Scanning Operation: When a valid channel list is defined,
INITiate[:IMMediate] begins the scan and closes the first channel in
the channel list. Successive triggers from the source specified by
TRIGger:SOURce advance the scan through the channel list.
•
Stopping Scan: See ABORt.
•
Related Commands: TRIGger, TRIGger:SOURce
•
*RST Condition: All channels open.
Example
Scanning Using External Device:
See Chapter 4, "Example: Scanning With External Devices" or "Example:
Scanning Using Trig Out and Event In Ports" for example scanning
programs using external instruments.
54 Agilent E1364A Form C Switch Command Reference
Chapter 5
Содержание E1364A
Страница 2: ......
Страница 6: ...Notes 4 Agilent E1364A Switch Module Table of Contents ...
Страница 10: ...Notes 8 Agilent E1364A 16 Channel Form C Switch User s Manual ...
Страница 11: ...Notes Agilent E1364A 16 Channel Form C Switch User s Manual 9 ...
Страница 12: ...Notes 10 Agilent E1364A 16 Channel Form C Switch User s Manual ...
Страница 18: ...16 Getting Started with the Agilent E1364A Chapter 1 ...
Страница 21: ...Figure 2 2 Wiring the Form C Switch Terminal Block Chapter 2 Configuring the Agilent E1364A Form C Switch 19 ...
Страница 70: ...68 Agilent E1364A Form C Switch Registers Appendix B ...
Страница 72: ...70 Agilent E1364A Form C Switch Error Messages Appendix C ...
Страница 76: ...74 Agilent E1364A 16 Channel Form C Switch Index ...