252
Chapter 5 Tutorial
Burst
5
For bursts, the trigger source can be an external signal, an internal
timer, the
key, or a command received from the remote interface.
The input for external trigger signals is the rear-panel
Ext Trig
connector. This connector accepts TTL-compatible levels and is referenced
to chassis ground (not floating ground). When not used as an input, the
Ext Trig
connector can be configured as an output to enable the 33500
Series to trigger other instruments at the same time that its internal
trigger occurs.
An N-Cycle burst always begins and ends at the same point in the
waveform, called the
start phase
.
Gated Burst
In the Gated burst mode, the output waveform is either
“on” or “off ” based on the level of the external signal applied to the rear-
panel
Ext Trig
connector. When the gate signal is
true
, the waveform
generator outputs a continuous waveform. When the gate signal goes
false
, the current waveform cycle is completed and then the waveform
generator stops while remaining at the voltage level corresponding to the
start phase of the selected waveform. For a noise waveform, the output
stops immediately when the gate signal goes false.
Содержание Agilent 33500 Series
Страница 2: ...Agilent Technologies Agilent 33500 Series 30 MHz Function Arbitrary Waveform Generator User s Guide ...
Страница 13: ...12 ...
Страница 16: ...1 1 Quick Start ...
Страница 41: ...40 Chapter 1 Quick Start To Rack Mount the Waveform Generator 1 ...
Страница 42: ...2 2 Front Panel Menu Operation ...
Страница 90: ...3 3 Features and Functions ...
Страница 218: ...4 4 Application Programs ...
Страница 243: ...242 Chapter 4 Application Programs Program Listings 4 ...
Страница 244: ...5 5 Tutorial ...
Страница 261: ...260 Chapter 5 Tutorial Ground Loops 5 ...
Страница 262: ...6 Specifications ...