Chapter 3 Features and Functions
Burst Mode
3
Burst Mode
You can configure the waveform generator to output a waveform with a
specified number of cycles, called a
burst
. The waveform generator can
produce a burst using sine, square, triangle, ramp, pulse, PRBS, or
arbitrary waveforms (noise is allowed only in the gated burst mode and
DC is not allowed).
For more information on the fundamentals of the burst mode, refer to
chapter 5, “Tutorial.”
To Select Burst
• The waveform generator will not allow burst to be enabled at the
same time that sweep or any modulation mode is enabled. When you
enable burst, the sweep, list, or modulation mode is turned off.
•
Front-Panel Operation
: You must enable burst
before
setting up any
of the other burst parameters. Press
Burst On/Off
to output a burst
using the present settings for frequency, output amplitude, and offset
voltage.
•
Remote Interface Operation
: To avoid multiple waveform changes,
enable the burst mode
after
you have set up the other parameters.
[SOURce[1|2]:]BURSt:STATe {OFF|ON}
Содержание Agilent 33500 Series
Страница 2: ...Agilent Technologies Agilent 33500 Series 30 MHz Function Arbitrary Waveform Generator User s Guide ...
Страница 13: ...12 ...
Страница 16: ...1 1 Quick Start ...
Страница 41: ...40 Chapter 1 Quick Start To Rack Mount the Waveform Generator 1 ...
Страница 42: ...2 2 Front Panel Menu Operation ...
Страница 90: ...3 3 Features and Functions ...
Страница 218: ...4 4 Application Programs ...
Страница 243: ...242 Chapter 4 Application Programs Program Listings 4 ...
Страница 244: ...5 5 Tutorial ...
Страница 261: ...260 Chapter 5 Tutorial Ground Loops 5 ...
Страница 262: ...6 Specifications ...