Chapter 3 Features and Functions
Frequency Modulation (FM)
3
Modulating Waveform Shape
The waveform generator will accept an internal or external modulation
source for FM.
• Modulating waveform shape (
internal source
): Sine (default), Square,
Pulse, UpRamp, DnRamp, Triangle, Noise, or PRBS
.
• Square has 50% duty cycle
• UpRamp has 100% symmetry
• Triangle has 50% symmetry
• DnRamp has 0% symmetry
• Noise - white gaussian noise
• PRBS - Pseudo Random Bit Sequence (polynomial PN7)
• You can use noise as the modulating waveshape, but you
cannot
use
noise, arbitrary waveforms, or
DC
as the carrier waveform.
•
Front-Panel Operation
: After selecting FM, press the
Shape
softkey.
•
Remote Interface Operation
:
FM:INTernal:FUNCtion
{SINusoid|SQUare|RAMP|NRAMp|TRIangle|
NOISe|PRBS|ARB}
Содержание Agilent 33500 Series
Страница 2: ...Agilent Technologies Agilent 33500 Series 30 MHz Function Arbitrary Waveform Generator User s Guide ...
Страница 13: ...12 ...
Страница 16: ...1 1 Quick Start ...
Страница 41: ...40 Chapter 1 Quick Start To Rack Mount the Waveform Generator 1 ...
Страница 42: ...2 2 Front Panel Menu Operation ...
Страница 90: ...3 3 Features and Functions ...
Страница 218: ...4 4 Application Programs ...
Страница 243: ...242 Chapter 4 Application Programs Program Listings 4 ...
Страница 244: ...5 5 Tutorial ...
Страница 261: ...260 Chapter 5 Tutorial Ground Loops 5 ...
Страница 262: ...6 Specifications ...