![Agilent Technologies Agilent 33500 Series Скачать руководство пользователя страница 166](http://html.mh-extra.com/html/agilent-technologies/agilent-33500-series/agilent-33500-series_user-manual_2867489166.webp)
165
Chapter 3 Features and Functions
Triggering
3
Trigger Input Signal
This rear-panel connector is used in the following modes:
•
Triggered Sweep Mode
: To select the external source, press the
Trigger Setup
softkey and then select the
Source Ext
softkey or
execute the
TRIG:SOUR EXT
command from the remote interface
(sweep must be enabled). When the rising or falling edge (you specify
which edge) of a TTL pulse is received on the
Ext Trig
connector,
the waveform generator outputs a single sweep.
•
Externally-Modulated FSK Mode
: To enable the externally-modulated
mode, press the
Source
softkey from the front-panel or execute the
FSK:SOUR EXT
command from the remote interface (FSK must be
enabled). When a logic
low
level is present, the
carrier
frequency is
output. When a logic
high
level is present, the
hop
frequency is
output. The maximum external FSK rate is 100 kHz.
•
Triggered Burst Mode
: To select the external source, press the
Trigger Setup
softkey and then select the
Source Ext
softkey or
execute the
TRIG:SOUR EXT
command from the remote interface
(burst must be enabled). The waveform generator outputs a waveform
with a specified number of cycles (
burst count
) each time a trigger is
received from the specified trigger source.
•
External Gated Burst Mode
: To enable the gated mode, press the
Gated
softkey or execute
BURS:MODE GAT
from the remote interface
with burst enabled. When the external gate signal is
true
, the
waveform generator outputs a continuous waveform. When the
external gate signal goes
false
, the current waveform cycle is
completed and then the waveform generator stops while remaining at
the voltage level corresponding to the starting burst phase. For noise,
the output stops immediately when the gate signal goes false.
Содержание Agilent 33500 Series
Страница 2: ...Agilent Technologies Agilent 33500 Series 30 MHz Function Arbitrary Waveform Generator User s Guide ...
Страница 13: ...12 ...
Страница 16: ...1 1 Quick Start ...
Страница 41: ...40 Chapter 1 Quick Start To Rack Mount the Waveform Generator 1 ...
Страница 42: ...2 2 Front Panel Menu Operation ...
Страница 90: ...3 3 Features and Functions ...
Страница 218: ...4 4 Application Programs ...
Страница 243: ...242 Chapter 4 Application Programs Program Listings 4 ...
Страница 244: ...5 5 Tutorial ...
Страница 261: ...260 Chapter 5 Tutorial Ground Loops 5 ...
Страница 262: ...6 Specifications ...