![Agilent Technologies Agilent 33500 Series Скачать руководство пользователя страница 106](http://html.mh-extra.com/html/agilent-technologies/agilent-33500-series/agilent-33500-series_user-manual_2867489106.webp)
105
Chapter 3 Features and Functions
Output Configuration
3
Voltage Autoranging
Autoranging is enabled by default and the waveform generator
automatically selects the optimal settings for the attenuators. With
autoranging disabled, the waveform generator uses the current attenuator
settings and does not switch attenuator relays.
• You can disable autoranging to eliminate momentary disruptions
caused by switching of the attenuators while changing amplitude.
However, turning autoranging off has side effects:
• The amplitude and offset accuracy and resolution (and waveform
fidelity) may be adversely affected when reducing the amplitude
below a range change that would occur with autoranging on.
• You may not be able to achieve the minimum amplitude that is
available with autoranging on.
• Some instrument specifications do not apply with autoranging off.
•
Front-Panel Operation
: Press
or or
and press the
Range
softkey again to toggle between the
Auto
and
Hold
selections.
•
Remote Interface Operation
:
VOLTage:RANGe:AUTO {OFF|ON|ONCE
}
The
APPLy
command overrides the voltage autorange setting and
automatically enables autoranging.
Output Control
You can disable or enable the front-panel channel output connector. By
default, the output is disabled at power on to protect other equipment.
When a channel output connector is enabled, the corresponding
or
or
key is illuminated.
•
Front-Panel Operation
: Press
or
or
followed by
Output On/Off
to enable the output.
• If an excessive external voltage is applied to a front-panel channel
output connector by an external circuit, the instrument will generate
an error message and disable the output. To re-enable the output,
remove the overload from the connector and press
or
or
followed by
Output On/Off
to enable the output.
Содержание Agilent 33500 Series
Страница 2: ...Agilent Technologies Agilent 33500 Series 30 MHz Function Arbitrary Waveform Generator User s Guide ...
Страница 13: ...12 ...
Страница 16: ...1 1 Quick Start ...
Страница 41: ...40 Chapter 1 Quick Start To Rack Mount the Waveform Generator 1 ...
Страница 42: ...2 2 Front Panel Menu Operation ...
Страница 90: ...3 3 Features and Functions ...
Страница 218: ...4 4 Application Programs ...
Страница 243: ...242 Chapter 4 Application Programs Program Listings 4 ...
Страница 244: ...5 5 Tutorial ...
Страница 261: ...260 Chapter 5 Tutorial Ground Loops 5 ...
Страница 262: ...6 Specifications ...