Burst Mode
3
Burst Mode
You can use burst in one of two modes as described below. The selected
mode controls the allowable trigger source that you selected, and
whether certain other burst parameters apply (see the table below).
•
Triggered Burst Mode
: In this mode (default), the waveform
generator outputs a waveform with a specified number of cycles
(
burst count
) each time a trigger is received. After the specified
number of cycles have been output, the waveform generator stops and
waits for the next trigger. You can configure the waveform generator
to use an internal trigger to initiate the burst or you can provide an
external trigger by pressing the front-panel
key, applying a
trigger signal to the rear-panel
Ext Trig
connector, or sending a
software trigger command from the remote interface.
•
External Gated Burst Mode
: In this mode, the output waveform is
either “on” or “off ” based on the level of the external signal applied
to the rear-panel
Ext Trig
connector. When the gate signal is
true
,
the waveform generator outputs a continuous waveform. When the
gate signal goes
false
, the current waveform cycle is completed and
then the waveform generator stops while remaining at the voltage
level corresponding to the starting burst phase of the selected
waveform. For a noise waveform, the output stops immediately when
the gate signal goes false.
Burst Mode
(BURS:MODE)
Burst Count
(BURS:NCYC)
Burst Period
(BURS:INT:PER)
Burst Phase
(BURS:PHAS)
Trigger Source
(TRIG:SOUR)
Triggered Burst Mode:
Internal Trigger
TRIGgered
Available
Available
Available
IMMediate
Triggered Burst Mode:
External Trigger
TRIGgered
Available
Not Used
Available
EXTernal, BUS
Gated Burst Mode:
External Trigger
GATed
Not Used
Not Used
Available
Not Used
Timer Burst Mode:
Internal Trigger
TRIGgered
Available
Not Used
Available
TIMer
Содержание Agilent 33500 Series
Страница 2: ...Agilent Technologies Agilent 33500 Series 30 MHz Function Arbitrary Waveform Generator User s Guide ...
Страница 13: ...12 ...
Страница 16: ...1 1 Quick Start ...
Страница 41: ...40 Chapter 1 Quick Start To Rack Mount the Waveform Generator 1 ...
Страница 42: ...2 2 Front Panel Menu Operation ...
Страница 90: ...3 3 Features and Functions ...
Страница 218: ...4 4 Application Programs ...
Страница 243: ...242 Chapter 4 Application Programs Program Listings 4 ...
Страница 244: ...5 5 Tutorial ...
Страница 261: ...260 Chapter 5 Tutorial Ground Loops 5 ...
Страница 262: ...6 Specifications ...