Chapter 3 Features and Functions
Pulse Waveforms
3
Pulse Width
The pulse width represents the time from the 50% threshold of the
rising edge of the pulse to the 50% threshold of the next falling edge.
• Pulse width: 16 ns to 1,000,000 s (see restrictions below).
The default
pulse width is
100 µs.
• The specified pulse width must also be less than the difference
between the
period
and the
minimum pulse width as shown below
.
The waveform generator will adjust the pulse width as needed to
accommodate the specified period.
Pulse Width < Period – 16 ns
• The waveform generator will automatically adjust the pulse width as
needed to accommodate the specified period.
•
Front-Panel Operation
: After selecting the pulse function, press the
Pulse Width
softkey. Then use the knob or numeric keypad to enter
the desired pulse width.
•
Remote Interface Operation
:
FUNCtion:PULSe:WIDTh {<
seconds
>|MINimum|MAXimum}
Содержание Agilent 33500 Series
Страница 2: ...Agilent Technologies Agilent 33500 Series 30 MHz Function Arbitrary Waveform Generator User s Guide ...
Страница 13: ...12 ...
Страница 16: ...1 1 Quick Start ...
Страница 41: ...40 Chapter 1 Quick Start To Rack Mount the Waveform Generator 1 ...
Страница 42: ...2 2 Front Panel Menu Operation ...
Страница 90: ...3 3 Features and Functions ...
Страница 218: ...4 4 Application Programs ...
Страница 243: ...242 Chapter 4 Application Programs Program Listings 4 ...
Страница 244: ...5 5 Tutorial ...
Страница 261: ...260 Chapter 5 Tutorial Ground Loops 5 ...
Страница 262: ...6 Specifications ...