Error Messages
Attenuator Modules
414
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
Attenuator Modules
The following Error Codes are applicable to attenuator modules.
Codes E10010, E10012, and E10014 are valid for all attenuators. The remaining
codes are only valid for attenuators with power control.
error message
E10010: Invalid Caldata
description
The Caldata loaded on the module is invalid. Either the caldata ver-
sion number is not compatible with the firmware or the caldata is
corrupt (checksum incorrect)
test value
(no test value)
action
Please return the unit to your local service center for repair
error message
E10012: Dual ported RAM write/read test failed
description
Writing bit patterns and reading them back to/from the dual ported
ram failed
test value
(no test value)
action
Please return the unit to your local service center for repair
error message
E10013: FPGA not correctly initialized
description
Loading the FPGA with data failed. This indicates a defective FPGA
or an FPGA access problem
test value
(no test value)
action
Please return the unit to your local service center for repair
error message
E10014: Firmware corrupt, checksum error (a,b)
description
Firmware in flash has an invalid checksum. Either the image which
was programmed was corrupt or the flash memory has a failure
test value
a
is the stored checksum
b
is the calculated checksum
action
Reload the module firmware. If the failure persists, please return the
unit to your local service center for repair
error message
E10031: 0.25 ref out of limits (a, b, c)
description
The 0.25V reference voltage measured through the 16bit ADC is out
of the acceptable limits
test value
a
shows the ADC reading (0.25V = 3276)
b
shows the lower limit
c
shows the upper limit
action
Please return the unit to your local service center for repair
Содержание 8163B
Страница 4: ......
Страница 20: ...20 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 22: ...22 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 62: ...Getting Started A Sample Session 62 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 82: ...Additional Information How to Connect a Printer 82 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 108: ...Power Measurement How to Measure Power 108 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 118: ...Laser Sources How to Use Laser Source Modules 118 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 158: ...Tunable Lasers How to Use Auxiliary Functions 158 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 318: ...Installation and Maintenance Claims and Repackaging 318 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 334: ...Accessories GPIB Cables and Adapters 334 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 368: ...Performance Tests Test Record 368 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 378: ...Performance Tests Test Record 378 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 450: ...Cleaning Instructions Other Cleaning Hints 450 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 462: ...Firmware Upgrades Firmware Upgrade Process 462 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 465: ......