Return Loss Measurement
Return Loss Measurement
180
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
Calibration using the Agilent 81610CC
Reference Cable
Use the Agilent 816100CC Refernce Cable to make a Reflectance
Calibration, and the measurement patchcord to make a Termination
Calibration.
To measure insertion loss and the front panel delta of the system (see
“
Calculating the Front Panel Delta
” on page 195), you must also measure
the insertion loss of the Reference Cable and mesurement patchcord.
Reflectance Calibration
Use this procedure to calibrate the Return Loss module against a
component of known return loss.
1
Making sure all the connectors are clean, set up the instrument as shown in
Figure 98 if you are using an External Source,
Figure 98 Measuring the Return Loss of the Reference Cable - External Source
or Figure 99 if you are using an internal source
(Agilent 81611A/2A/3A/4A Return Loss modules only).
Figure 99 Measuring the Return Loss of the Reference Cable - Internal Source
Attach the high return loss connector of the reference cable to the output.
Tip: For best results and improved repeatability, fix the cable.
2
Move to [FPDelta], press [Edit], set the value to 0.000, then press [OK].
Agilent 81610CC Reference Cable
8161x
8163B Lightwave Multimeter
Laser
Source
Agilent 81113PC
8161x
8163B Lightwave Multimeter
Agilent 81610CC Reference Cable
Содержание 8163B
Страница 4: ......
Страница 20: ...20 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 22: ...22 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 62: ...Getting Started A Sample Session 62 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 82: ...Additional Information How to Connect a Printer 82 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 108: ...Power Measurement How to Measure Power 108 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 118: ...Laser Sources How to Use Laser Source Modules 118 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 158: ...Tunable Lasers How to Use Auxiliary Functions 158 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 318: ...Installation and Maintenance Claims and Repackaging 318 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 334: ...Accessories GPIB Cables and Adapters 334 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 368: ...Performance Tests Test Record 368 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 378: ...Performance Tests Test Record 378 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 450: ...Cleaning Instructions Other Cleaning Hints 450 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 462: ...Firmware Upgrades Firmware Upgrade Process 462 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 465: ......