The Logging Application
Applications
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
247
13
Move to <Ref Mode>, press Enter, move to one of the following referencing
modes:
•
<Value>, you set a reference value using <Ref>, see step 14,
•
<First Sample>, the first sample for each channel of the logging function is
selected as the reference value for all subsequent measurements for that
channel,
•
<Channel 1>, the value that Channel 1 measures is updated as the reference
value for each set of power measurements (thus, each data value for
channel 1 will be 0 dB), and
•
press Enter.
14
If you have chosen <Value> as the <Ref Mode>, move to <Ref>, press the
[Edit] softkey, enter a value for the reference, and press Enter.
Setting Logging Parameter
15
Move to [Samples], press Enter, enter an integer value between 1 and 4000, and
press Enter.
16
Move to <AvgTime>, press Enter, move to an averaging time period, and press
Enter.
These settings only have an effect when you choose <dB> as the
<Pwr unit>. See step 8 or step 3.
N O T E
Содержание 8163B
Страница 4: ......
Страница 20: ...20 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 22: ...22 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 62: ...Getting Started A Sample Session 62 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 82: ...Additional Information How to Connect a Printer 82 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 108: ...Power Measurement How to Measure Power 108 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 118: ...Laser Sources How to Use Laser Source Modules 118 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 158: ...Tunable Lasers How to Use Auxiliary Functions 158 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 318: ...Installation and Maintenance Claims and Repackaging 318 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 334: ...Accessories GPIB Cables and Adapters 334 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 368: ...Performance Tests Test Record 368 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 378: ...Performance Tests Test Record 378 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 450: ...Cleaning Instructions Other Cleaning Hints 450 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 462: ...Firmware Upgrades Firmware Upgrade Process 462 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 465: ......